UNIFORM IRRADIATION OF THE CRANIOSPINAL AXIS WITH A PENUMBRA MODIFIER AND AN ASYMMETRIC COLLIMATOR

被引:8
作者
SOHN, JW
SCHELL, MC
DASS, KK
SUH, JH
TEFFT, M
机构
[1] Cleveland Clinic Foundation, Department of Radiation Oncology, Cleveland
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1994年 / 29卷 / 01期
关键词
CRANIOSPINAL IRRADIATION; ASYMMETRIC COLLIMATOR; PENUMBRA MODIFIER;
D O I
10.1016/0360-3016(94)90243-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: A technique is described that uses an independent, asymmetric collimator and a penumbra modifier to uniformly administer radiation over the craniospinal axis. Method and Materials: An isocenter is used at the junction of the cranial and the upper spinal fields. These fields are defined by a single isocenter at spinal cord depth and an independent, asymmetric collimator. From the isocenter, the cranial and upper spinal fields are extended 1 cm inferiorly and superiorly, respectively. A modifier provides a 2-cm wide penumbra at the central axis of the beam (overlapping region) and is attached to a wedge tray in an accessory slot. This modifier allows the fields to be matched so a uniform dose is delivered over the isocenter and junction. Dose distribution was measured with an anthropomorphic head-and-neck wax phantom that included the seventh cervical vertebrae. A film was placed in a coronal cut at the spinal cord level. Results: The administered dose varied less than 10% through the craniospinal axis. Reproducibility with portal films has been very good. Advantages include dose homogeneity, easy reproducibility, and decreased setup time. Conclusion: This single-field isocentric technique allows more uniform irradiation of the craniospinal axis than do previously described techniques.
引用
收藏
页码:187 / 190
页数:4
相关论文
共 7 条
[1]   A MODIFIED RADIOTHERAPY TECHNIQUE IN THE TREATMENT OF MEDULLOBLASTOMA [J].
DEWIT, L ;
VANDAM, J ;
RIJNDERS, A ;
VANDEVELDE, G ;
ANG, KK ;
VANDERSCHUEREN, E .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1984, 10 (02) :231-241
[2]   CLINICAL USE OF A MATCH-LINE WEDGE FOR ADJACENT MEGAVOLTAGE RADIATION-FIELD MATCHING [J].
FRAASS, BA ;
TEPPER, JE ;
GLATSTEIN, E ;
VANDEGEIJN, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (02) :209-216
[3]   THE DOSIMETRY OF A SINGLE HOCKEY STICK PORTAL FOR TREATMENT OF TUMORS OF THE CRANIO-SPINAL AXIS [J].
GLASGOW, GP ;
MARKS, JE .
MEDICAL PHYSICS, 1983, 10 (05) :672-675
[4]  
GRIFFIN TW, 1976, BRIT J RADIOL, V49, P888
[5]   FIELD MATCHING CONSIDERATIONS IN CRANIOSPINAL IRRADIATION [J].
TATCHER, M ;
GLICKSMAN, AS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (04) :865-869
[6]   MEDULLOBLASTOMA - TREATMENT TECHNIQUE AND RADIATION-DOSIMETRY [J].
VANDYK, J ;
JENKIN, RDT ;
LEUNG, PMK ;
CUNNINGHAM, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1977, 2 (9-10) :993-1005
[7]   TECHNIQUE FOR MATCHING ORTHOGONAL MEGAVOLTAGE FIELDS [J].
WILLIAMSON, TJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1979, 5 (01) :111-116