APPLICATION OF A SURFACE-ACOUSTIC-WAVE DEVICE FOR MEASUREMENT OF LIQUID FLOW-RATE

被引:4
作者
JOSHI, SG
JIN, Y
机构
[1] Department of Electrical and Computer Engineering, Marquette University, Milwaukee, WI 53233
关键词
D O I
10.1109/58.105254
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The use of a surface-acoustic-wave (SAW) device to measure the flow rate of liquids is described. A delay-line stabilized SAW oscillator heated to a suitable temperature above ambient is cooled by the flowing fluid. This results in a change in the oscillator frequency. The frequency of a 68-MHz oscillator operated at 9°C above ambient is found to vary by more than 40 kHz for variation in water flow rate from 0. to 0.8 ml/min. © 1990 IEEE