CLINICAL AND LABORATORY EVALUATION OF AN ELECTROSURGICAL LAPAROSCOPIC TROCAR

被引:10
作者
WAXMAN, K
BIRKETT, DH
SACKIER, JM
ESTEMCDONALD, J
DUQUETTE, J
机构
[1] Department of Surgery, University of California Irvine Medical Center, Orange, 92668, CA
[2] Boston University Medical Center, Boston, 02118, MA
[3] University of California, San Diego, San Diego, 92103, CA
来源
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES | 1994年 / 8卷 / 09期
关键词
LAPAROSCOPY; TROCAR; TROCAR INJURY; COMPLICATION;
D O I
10.1007/BF00705723
中图分类号
R61 [外科手术学];
学科分类号
摘要
Electrosurgical energy may be utilized as an adjunct to mechanical force for insertion of laparoscopic trocars. The advantage of this approach may be better operator control of insertion, with less risk of intraperitoneal and retroperitoneal injury. To assess the safety and efficacy of electrosurgical trocars, we compared them to mechanical trocars in clinical and animal trials. During 100 trocar introductions in 25 laparoscopic cholecystectomies, insertion force was measured. In contrast to mechanical torcars, which required progressively more force to insert as size increased, electrosurgical trocars required the same low insertion force regardless of size. No wound complications occurred. In animal experiments, wound healing (measured histologically and by bursting strength) was normal and equivalent for mechanical and electrosurgical insertions. We conclude that electrosurgical trocars require less force for insertion and do not impair would healing. Electrosurgical trocars may thus offer important safety advantages over mechanical trocars.
引用
收藏
页码:1076 / 1079
页数:4
相关论文
共 5 条
[1]  
Baadsgaard S.E., Bille S., Egeblad K., Major vascular injury during gynecologic laparoscopy. Report of a case and review of published cases, Acta Obstet Gynecol Scand, 68, pp. 283-285, (1989)
[2]  
Chapron C., Querleu D., Mage G., Madelenat P., Dubuisson J.B., Audebert A., Erny R., Bruhat M.A., Complications of gynecologic laparoscopy. Multicentric study of 7604 laparoscopies, J Gynecol Obst Bio Reprod, 21, pp. 207-213, (1992)
[3]  
Cunanan R.G., Courey N.G., Lippes J., Complications of laparoscopic tubal sterilization, Obstet Gynecol, 55, pp. 501-506, (1980)
[4]  
Lightdale C., Indications, contraindications and complications of laparoscopy, Gastroenterologic endoscopy, pp. 1039-1044, (1987)
[5]  
Peterson H.B., Greenspan J.R., Ory W.H., Death after puncture of the aorta during laparoscopic sterilization, Am J Obstet Gynecolol, 59, pp. 133-134, (1982)