CHEST RADIOGRAPHS OBTAINED WITH SHAPED FILTERS - EVALUATION BY OBSERVER PERFORMANCE TESTS

被引:9
作者
KELSEY, CA
LANE, RG
MOSELEY, RD
METTLER, FA
ROSENBERG, RD
WILLIAMS, AG
GARCIA, JF
FELDMAN, BS
BOARDMAN, RE
机构
关键词
D O I
10.1148/radiology.159.3.3704147
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The effectiveness of a shaped filter in improving nodule and infiltrate detection was measured by observer performance testing. Seven observers read 152 test radiographs of the chest obtained from human volunteers. Half the test radiographs had target images. Observer performance in detecting nodule or infiltrate images was compared with the shaped-filter system and with a conventional chest imaging system. The results were analyzed using receiver operating characteristic (ROC) techniques and indicate that the filter technique was not significantly different from the conventional technique in infiltrate depiction. Observer performance in detecting nodules was slightly worse on images obtained with the shaped-filter system.
引用
收藏
页码:653 / 655
页数:3
相关论文
共 11 条
[1]  
[Anonymous], 1975, NORMALITY PREDICTIVE, P131
[2]   SHAPED, LEAD-LOADED ACRYLIC FILTERS FOR PATIENT EXPOSURE REDUCTION AND IMAGE-QUALITY IMPROVEMENT [J].
GRAY, JE ;
STEARS, JG ;
FRANK, ED .
RADIOLOGY, 1983, 146 (03) :825-828
[3]   LUNG-CANCER DETECTED DURING A SCREENING-PROGRAM USING 4-MONTH CHEST RADIOGRAPHS [J].
MUHM, JR ;
MILLER, WE ;
FONTANA, RS ;
SANDERSON, DR ;
UHLENHOPP, MA .
RADIOLOGY, 1983, 148 (03) :609-615
[4]  
NAIMUDDIN S, 1984, Radiology, V153, P38
[5]  
NAIMUDDIN S, 1984, MED PHYS, V11, P364
[6]  
Plewes D. B., 1980, Proceedings of the Society of Photo-Optical Instrumentation Engineers, V233, P183
[7]  
Sorenson J A, 1981, Invest Radiol, V16, P529, DOI 10.1097/00004424-198111000-00015
[8]   PHOTOGRAPHIC UNSHARP MASKING IN CHEST RADIOGRAPHY [J].
SORENSON, JA ;
NIKLASON, LT ;
NELSON, JA .
INVESTIGATIVE RADIOLOGY, 1981, 16 (04) :281-288
[9]  
VYBORNY C, 1984, RADIOLOGY, V151, P154
[10]   IMPROVED ROUTINE CHEST RADIOGRAPHY WITH A TROUGH FILTER [J].
WIEDER, S ;
ADAMS, PL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 137 (04) :695-698