1ST COMPARISON OF THE SOLAR AND AN SI RADIOMETRIC SCALE

被引:21
作者
ROMERO, J
FOX, NP
FROHLICH, C
机构
[1] DAVOS WORLD RADIAT CTR, PHYS METEOROL OBSERV, CH-7260 DAVOS, SWITZERLAND
[2] NATL PHYS LAB, DIV QUANTUM METROL, TEDDINGTON TW11 0LW, MIDDX, ENGLAND
关键词
D O I
10.1088/0026-1394/28/3/004
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The paper presents the results of the first comparison between the solar radiometric scale, the World Radiometric Reference (WRR) and a Systeme International radiometric scale (SI scale) as disseminated by national standards laboratories, in particular, the scale based on the cryogenic radiometer of the National Physical Laboratory (NPL). The comparison used a number of transfer standards to act as intermediaries between the two scales which limited the overall uncertainty to 0,07%. The final result shows that the WRR and the NPL SI scale agree to better than 0,03%, well within the absolute uncertainty of the two scales.
引用
收藏
页码:125 / 128
页数:4
相关论文
共 11 条
[1]   ABSOLUTE RADIOMETERS (PMO6) AND THEIR EXPERIMENTAL CHARACTERIZATION [J].
BRUSA, RW ;
FROHLICH, C .
APPLIED OPTICS, 1986, 25 (22) :4173-4180
[2]  
FOUKAL P, 1983, P SOC PHOTO-OPT INST, V416, P197, DOI 10.1117/12.935936
[3]   COMPARISON OF 2 CRYOGENIC RADIOMETERS BY DETERMINING THE ABSOLUTE SPECTRAL RESPONSIVITY OF SILICON PHOTODIODES WITH AN UNCERTAINTY OF 0.02-PERCENT [J].
FOX, NP ;
MARTIN, JE .
APPLIED OPTICS, 1990, 29 (31) :4686-4693
[4]   TRAP DETECTORS AND THEIR PROPERTIES [J].
FOX, NP .
METROLOGIA, 1991, 28 (03) :197-202
[5]   HISTORY OF SOLAR RADIOMETRY AND THE WORLD RADIOMETRIC REFERENCE [J].
FROHLICH, C .
METROLOGIA, 1991, 28 (03) :111-115
[6]  
FROHLICH C, 1978, WMO FINAL REPORT, V490, P108
[7]   A CRYOGENIC RADIOMETER FOR ABSOLUTE RADIOMETRIC MEASUREMENTS [J].
MARTIN, JE ;
FOX, NP ;
KEY, PJ .
METROLOGIA, 1985, 21 (03) :147-155
[8]   A RADIOMETRIC DETERMINATION OF THE STEFAN-BOLTZMANN CONSTANT AND THERMODYNAMIC TEMPERATURES BETWEEN -40-DEGREES-C AND + 100-DEGREES-C [J].
QUINN, TJ ;
MARTIN, JE .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 1985, 316 (1536) :85-+
[9]   OPTICAL POWER CALIBRATOR BASED ON A STABILIZED GREEN HE-NE-LASER AND A CRYOGENIC ABSOLUTE RADIOMETER [J].
VARPULA, T ;
SEPPA, H ;
SAARI, JM .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1989, 38 (02) :558-564
[10]  
ZALEWSKI E, 1986, BIPM COM CONS PHOT R, V11, P165