DEPTH OF TISSUE SAMPLING IN THE OPTIC-NERVE HEAD USING LASER DOPPLER FLOWMETRY

被引:72
作者
KOELLE, JS
RIVA, CE
PETRIG, BL
CRANSTOUN, SD
机构
[1] UNIV PENN,SCHEIE EYE INST,51 N 39TH ST,PHILADELPHIA,PA 19104
[2] UNIV PENN,DEPT OPHTHALMOL,PHILADELPHIA,PA 19104
关键词
LASER DOPPLER FLOWMETRY (LDF); SAMPLING DEPTH IN LDF; OPTIC NERVE BLOOD FLOW; LIGHT SCATTERING;
D O I
10.1007/BF02559756
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Laser Doppler flowmetry (LDF) was performed on a simulated blood vessel in a model eye through optic nerve tissue sections in order to ascertain the ability to detect flow through them. LDF was performed using either near-infra-red or green laser light. Tissue section thickness ranged from 50 μm to 1000 μm. As expected, we found that our ability to detect flow with LDF decreased as we increased the thickness of optic nerve sections interposed between the LDF apparatus and the simulated blood vessel. We also found that the sampled depth of LDF increased with increasing separation of the optical detection fibre from the centre of the illuminated tissue volume. With adequate separation, we were able to detect flow with LDF through tissue sections of 1000 μm thickness using either near-infra-red or green laser light. © 1993 Baillière Tindall.
引用
收藏
页码:49 / 54
页数:6
相关论文
共 6 条
[1]  
Bonner R., 1990, LASER DOPPLER BLOOD, P17
[2]  
HAYREH SS, 1989, OCULAR BLOOD FLOW GL, P3
[3]   PARTITIONAL MEASUREMENT OF CAPILLARY AND ARTERIOVENOUS ANASTOMOTIC BLOOD-FLOW IN THE HUMAN FINGER BY LASER-DOPPLER-FLOWMETER [J].
HIRATA, K ;
NAGASAKA, T ;
NODA, Y .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 1988, 57 (05) :616-621
[4]  
Riva C. E., 1989, OCULAR BLOOD FLOW GL, P129
[5]   FLICKER EVOKED INCREASE IN OPTIC-NERVE HEAD BLOOD-FLOW IN ANESTHETIZED CATS [J].
RIVA, CE ;
HARINO, S ;
SHONAT, RD ;
PETRIG, BL .
NEUROSCIENCE LETTERS, 1991, 128 (02) :291-296
[6]   FUNDUS CAMERA BASED RETINAL LDV [J].
RIVA, CE ;
GRUNWALD, JE ;
SINCLAIR, SH ;
OKEEFE, K .
APPLIED OPTICS, 1981, 20 (01) :117-120