A NEW DEVICE TO REMOVE OBSTRUCTION FROM ENDOTRACHEAL-TUBES DURING MECHANICAL VENTILATION IN CRITICALLY ILL PATIENTS

被引:12
作者
CONTI, G [1 ]
ROCCO, M [1 ]
DEBLASI, RA [1 ]
LAPPA, A [1 ]
ANTONELLI, M [1 ]
BUFI, M [1 ]
GASPARETTO, A [1 ]
机构
[1] UNIV ROMA LA SAPIENZA,INST ANESTHESIA & INTENS CARE,ROME,ITALY
关键词
MECHANICAL VENTILATION; ENDOTRACHEAL TUBES; OBSTRUCTION; ACUTE RESPIRATORY FAILURE;
D O I
10.1007/BF01705724
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To evaluate the efficiency of a new device developed to remove obstructions from endotracheal tubes (ETT) in mechanically ventilated patients. Design: Open study in mechanically ventilated sedated and paralyzed ICU patients. Setting: General ICU and Laboratory of Respiratory Mechanics of the University of Rome ''La Sapienza''. Patients: 8 consecutive unselected mechanically ventilated, critically ill patients in which a partial obstruction of ETT was suspected on the basis of an increase of the peak inspiratory pressure (>20%) plus the difficult introduction of a standard suction catheter. Interventions: Obstructions to ETT were removed with an experimental ''obstruction remover'' (OR) Measurements: ''In vive'' ETT airflow resistance (0.25; 0.5; 0.75; 1 l/s) was evaluated before and after use of the OR; the work of breathing necessary to overcome ETT resistance (WOBett) was also evaluated before and after OR use. Results: The use of OR significantly reduced in all patients the ETT ''in vive'' resistance (From 5.5+/-2.3 to 2.9+/-0.5 cmH(2)O/l/s at 0.25 l/s, p<0.05; from 9+/-2.4 to 3.8+/-0.8 cmH(2)O/l/s at 0.51/s; from 12.2,3.5 to 5.7+/-1.2 cmH(2)O/l/s at 0.75 l/s; from 16.9+/-6 to 9.3+/-3.8 cmH(2)O/l/s at 1 l/s, p<0.01 respectively). Also the WOBett was significantly reduced after use of the OR (from 0.66+/-0.19 to 0.34+/-0.08 J/1; p<0.05) Conclusion: this experimental device can be safely and successfully used to remove obstructions from the ETT lumen, without suspending mechanical ventilation, reducing the need for rapid ETT substitution in emergency and life-threatening situations.
引用
收藏
页码:573 / 576
页数:4
相关论文
共 9 条
[1]   IMPOSED WORK OF BREATHING AND METHODS OF TRIGGERING A DEMAND-FLOW, CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM [J].
BANNER, MJ ;
BLANCH, PB ;
KIRBY, RR .
CRITICAL CARE MEDICINE, 1993, 21 (02) :183-190
[2]  
CINNELLA G, 1989, REANIM SOINS INTENSI, V5, pA459
[3]   ENDOTRACHEAL-TUBE OCCLUSION ASSOCIATED WITH THE USE OF HEAT AND MOISTURE EXCHANGERS IN THE INTENSIVE-CARE UNIT [J].
COHEN, IL ;
WEINBERG, PF ;
FEIN, IA ;
ROWINSKI, GS .
CRITICAL CARE MEDICINE, 1988, 16 (03) :277-279
[4]   A FIBER OPTICS SYSTEM FOR THE EVALUATION OF AIRWAY PRESSURE IN MECHANICALLY VENTILATED PATIENTS [J].
DEBLASI, RA ;
CONTI, G ;
ANTONELLI, M ;
BUFI, M ;
GASPARETTO, A .
INTENSIVE CARE MEDICINE, 1992, 18 (07) :405-409
[5]   USE OF THE BLASIUS RESISTANCE FORMULA TO ESTIMATE THE EFFECTIVE DIAMETER OF ENDOTRACHEAL-TUBES [J].
LOFASO, F ;
LOUIS, B ;
BROCHARD, L ;
HARF, A ;
ISABEY, D .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1992, 146 (04) :974-979
[6]   BEDSIDE ESTIMATION OF THE INSPIRATORY WORK OF BREATHING DURING MECHANICAL VENTILATION [J].
MARINI, JJ ;
RODRIGUEZ, RM ;
LAMB, V .
CHEST, 1986, 89 (01) :56-63
[7]  
MARINI JJ, 1991, PULMONARY FUNCTION M, P62
[8]   COMPLICATIONS OF ASSISTED VENTILATION - PROSPECTIVE STUDY OF 354 CONSECUTIVE EPISODES [J].
ZWILLICH, CW ;
PIERSON, DJ ;
CREAGH, CE ;
SUTTON, FD ;
SCHATZ, E ;
PETTY, TL .
AMERICAN JOURNAL OF MEDICINE, 1974, 57 (02) :161-170
[9]  
1983, EMERG CARE RES I, V121, P155