MR ANGIOGRAPHY WITH A CARDIAC-PHASE - SPECIFIC ACQUISITION WINDOW

被引:26
作者
SELBY, K
SALONER, D
ANDERSON, CM
CHIEN, D
LEE, RE
机构
[1] Department of Radiology, University of California, San Francisco
[2] Radiology Services (114D), Veterans Administration Medical Center, San Francisco, California, 94121
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1992年 / 2卷 / 06期
关键词
ARTIFACT; EXTREMITIES; MR; FLOW SIMULATION; IMAGE PROCESSING; PHANTOMS; PULSE SEQUENCES; RAPID IMAGING; TECHNOLOGY; TISSUE SUPPRESSION; VASCULAR STUDIES;
D O I
10.1002/jmri.1880020606
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A method for cardiac-phase-specific magnetic resonance (MR) angiography is presented. An electronics module permits incrementing of phase-encoding gradients and storage of incoming data only during a chosen portion of the cardiac cycle. Suppression of stationary material is maintained by delivering radio-frequency pulses at constant TR throughout the cycle. Imaging of a pulsatile flow phantom demonstrates that acquiring data only during systole substantially increases the signal intensity of flowing material. In addition, phase-encoding ghost artifacts are eliminated from the neighborhood of the vessel. Image acquisition time is minimized by acquiring only the low-frequency phase-encoding lines in the cardiac-phase-specific mode. In healthy volunteers, greatly improved MR angiograms of the lower extremities are obtained. Fat saturation and magnetization transfer further enhance vessel/background contrast. Acquiring data only during systole ensures rapid inflow for all phase-encoding lines, permitting a near-longitudinal section orientation without in-plane saturation. This substantially reduces total acquisition time relative to axial acquisition.
引用
收藏
页码:637 / 643
页数:7
相关论文
共 10 条
[1]  
de Graaf RG, Groen JP, MR angiography with pulsatile flow, Magn Reson Imaging, 10, pp. 25-34, (1992)
[2]  
Perman WH, Moran PR, Moran RA, Bernstein MA, Artifacts from pulsatile flow in MR imaging, J Comput Assist Tomogr, 10, pp. 473-483, (1986)
[3]  
Teitelbaum GP, Ortega HV, Vinitski S, Et al., Optimization of gradient‐echo imaging parameters for intracaval filters and trapped thromboemboli, Radiology, 174, pp. 1013-1019, (1990)
[4]  
van Dijk P, Direct cardiac NMR imaging of heart wall and blood flow velocity, J Comput Assist Tomogr, 8, pp. 429-436, (1984)
[5]  
Firmin DN, Nayler GL, Klipstein RH, Underwood SR, Rees RSO, Longmore DB. In vivo validation of MR velocity imaging, J Comput Assist Tomogr, 11, pp. 751-756, (1987)
[6]  
Pelc NJ, Herfkens RJ, Shimakawa A, Enzmann DR, Phase contrast cine magnetic resonance imaging, Magn Reson Q, 7, pp. 229-254, (1991)
[7]  
Edelman RR, Ahn SS, Chien D, Et al., Improved time‐of‐flight MR angiography of the brain with magnetization transfer contrast, Radiology, 184, pp. 395-399, (1992)
[8]  
Pike GB, Hu BS, Glover GH, Enzmann DR, Magnetization transfer time‐of‐flight magnetic resonance angiography, Magn Reson Med, 25, pp. 372-379, (1992)
[9]  
Lanzer P, McKibbin W, Bohning D, Pohost G, Quantitation of abdominal aortic wall dynamics in man by gradient echo NMR imaging, Magn Reson Med, 7, pp. 407-415, (1990)
[10]  
Feinberg DA, Mark AS, Human brain motion and cerebrospinal fluid circulation demonstrated with MR velocity imaging, Radiology, 163, pp. 793-799, (1987)