COMPARISON OF 316LVM AND MP35N ALLOYS AS CHARGE INJECTION ELECTRODES

被引:10
作者
COGAN, SF [1 ]
JONES, GS [1 ]
HILLS, DV [1 ]
WALTER, JS [1 ]
RIEDY, LW [1 ]
机构
[1] EDWARD HINES VET ADM MED CTR,CTR REHABIL RES & DEV,HINES,IL 60141
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1994年 / 28卷 / 02期
关键词
D O I
10.1002/jbm.820280214
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An In vitro comparison of the corrosion response of 316LVM stainless steel and MP35N (a CoNiCrMo alloy) electrodes under conditions appropriate to applications in functional electrical stimulation (FES) was made. Electrodes of both alloys were subjected to a cathodic 40 mu C/cm(2) charge injection protocol and the potential transient response was recorded over a 96 h period. The transient responses were compared with potentiodynamic polarization data used to establish the quasiequilibrium response of the alloys in the carbonate and phosphate-buffered saline electrolyte used in the study. The MP35N electrodes exhibited extensive pitting corrosion during charge injection, whereas little corrosion was observed on 316LVM electrodes. An explanation for the susceptibility of MP35N to corrosion during charge injection is found in the potentiodynamic polarization data, which reveal a breakdown potential (critical pitting potential) of 0.45 V (SCE) for MP35N compared with 1.05 V (SCE) for 316LVM. Factors that may influence corrosion response during charge injection from alloys exhibiting active-passive behavior are discussed. (C) 1994 John Wiley and Sons, Inc.
引用
收藏
页码:233 / 240
页数:8
相关论文
共 20 条
[1]  
Agnew W. F., 1990, NEURAL PROSTHESES FU
[2]   CALCULATIONS OF THE PH CHANGES PRODUCED IN BODY TISSUE BY A SPHERICAL STIMULATION ELECTRODE [J].
BALLESTRASSE, CL ;
RUGGERI, RT ;
BECK, TR .
ANNALS OF BIOMEDICAL ENGINEERING, 1985, 13 (05) :405-424
[3]  
BENCE JL, 1986, BIOMEDICAL MATERIALS, P387
[4]   CONCEPT OF PROTECTION POTENTIAL APPLIED TO CORROSION OF METALLIC ORTHOPEDIC IMPLANTS [J].
CAHOON, JR ;
BANDYOPADHYA, R ;
TENNESE, L .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1975, 9 (03) :259-264
[5]   COMPARATIVE CREVICE CORROSION-RESISTANCE OF CO-CR BASE SURGICAL IMPLANT ALLOYS [J].
DEVINE, TM ;
WULFF, J .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1976, 123 (10) :1433-1437
[6]   COMPARATIVE CREVICE CORROSION-RESISTANCE OF CO-CR BASE SURGICAL IMPLANT ALLOYS [J].
DEVINE, TM ;
DEVINE, TM ;
WULFF, J .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1977, 124 (06) :869-870
[7]  
Gamble J.L., 1954, CHEM ANATOMY PHYSL P
[8]   DEVELOPMENT OF PERCUTANEOUS INTRAMUSCULAR ELECTRODE FOR MULTICHANNEL FES SYSTEM [J].
HANDA, Y ;
HOSHIMIYA, N ;
IGUCHI, Y ;
ODA, T .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (07) :705-710
[9]  
KRUGER J, 1984, 9TH P INT C MET CORR, V1, P79
[10]   PITTING CORROSION OF HIGH-STRENGTH ALLOY STIMULATION ELECTRODES UNDER DYNAMIC CONDITIONS [J].
LAN, N ;
DAROUX, M ;
MORTIMER, JT .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1989, 136 (04) :947-954