CONTROLLED CLINICAL-EVALUATION OF BACTEC PLUS AEROBIC/F AND BACT/ALERT AEROBIC FAN BOTTLES FOR DETECTION OF BLOOD-STREAM INFECTIONS

被引:57
作者
POHLMAN, JK
KIRKLEY, BA
EASLEY, KA
BASILLE, BA
WASHINGTON, JA
机构
[1] CLEVELAND CLIN FDN,DEPT CLIN PATHOL L40,CLEVELAND,OH 44195
[2] CLEVELAND CLIN FDN,DEPT BIOSTAT & EPIDEMIOL,CLEVELAND,OH 44195
关键词
D O I
10.1128/JCM.33.11.2856-2858.1995
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
A total of 7,190 blood culture sets were obtained from adult patients with a suspected bloodstream infection. A 20-ml sample of blood was distributed equally between the aerobic FAN bottle which was monitored in the BacT/Alert system and a Plus Aerobic/F bottle which was monitored in the BACTEC 9240 system, A total of 988 positive cultures were obtained from 483 patients; however, only 453 positive cultures from 173 patients met the criteria for volume (greater than or equal to 8 mi per bottle) and clinical significance on the basis of concurrent case review required for data analysis, There were 25 and 68 false positives from the FAN and Plus Aerobic/F bottles, respectively, There were no statistically significant differences between systems in the number of positive cultures or septic episodes by species; however, the total number of Enterobacteriaceae and Pseudomonas aeruginosa isolates combined was significantly greater in the FAN bottle (P = 0.04), Detection times did not differ significantly between systems for positive cultures; however, episodes of Staphylococcus aureus bacteremia were detected significantly more rapidly from the FAN bottle ( = 0.005), There was no significant difference between systems in the detection of bloodstream infections in patients receiving antibiotics at the time of blood culture.
引用
收藏
页码:2856 / 2858
页数:3
相关论文
共 13 条
  • [1] INCREASED DETECTION OF STAPHYLOCOCCAL BACTEREMIA USING AN ANTI-MICROBIAL REMOVAL DEVICE
    CARLSON, LG
    TENOVER, FC
    PLORDE, JJ
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1985, 84 (04) : 509 - 512
  • [2] ROUTINE EVALUATION OF BACTEC NR-16A AND NR-17A MEDIA
    COURCOL, RJ
    DUROCHER, AV
    ROUSSELDELVALLEZ, M
    FRUCHART, A
    MARTIN, GR
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1988, 26 (09) : 1619 - 1622
  • [3] DETECTION OF BACTEREMIA WITH THE BACTEC B-16 RESIN BLOOD CULTURE-MEDIUM
    DEGIROLAMI, PC
    EICHELBERGER, K
    SIEGEL, J
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1984, 81 (05) : 643 - 646
  • [4] DETECTION OF BACTEREMIA IN PATIENTS RECEIVING ANTI-MICROBIAL THERAPY - AN EVALUATION OF THE ANTI-MICROBIAL REMOVAL DEVICE AND B-16 MEDIUM
    DOERN, GV
    GANTZ, NM
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1983, 18 (01) : 43 - 48
  • [5] CDC DEFINITIONS FOR NOSOCOMIAL INFECTIONS, 1988
    GARNER, JS
    JARVIS, WR
    EMORI, TG
    HORAN, TC
    HUGHES, JM
    [J]. AMERICAN JOURNAL OF INFECTION CONTROL, 1988, 16 (03) : 128 - 140
  • [6] ILSTRUP DM, 1978, DETECTION SEPTICEMIA, P31
  • [7] CLINICAL COMPARISON OF ISOLATOR AND BACTEC-660 RESIN MEDIA FOR BLOOD CULTURE
    KELLY, MT
    ROBERTS, FJ
    HENRY, D
    GEERE, I
    SMITH, JA
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1990, 28 (09) : 1925 - 1927
  • [8] MULTICENTER COMPARISON OF THE HIGH-VOLUME (10 ML) NR BACTEC PLUS SYSTEM AND THE STANDARD (5 ML) (NR BACTEC SYSTEM
    KOONTZ, FP
    FLINT, KK
    REYNOLDS, JK
    ALLEN, SD
    [J]. DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 1991, 14 (02) : 111 - 118
  • [9] EVALUATION OF THE BACTEC ANTI-MICROBIAL REMOVAL SYSTEM FOR DETECTION OF BACTEREMIA
    MCGUIRE, NM
    KAUFFMAN, CA
    HERTZ, CS
    KOVACH, JM
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1983, 18 (03) : 449 - 451
  • [10] EVALUATION OF NEW BLOOD CULTURE PROCESSING SYSTEMS
    MOODY, JA
    FASCHING, CE
    SHANHOLTZER, CJ
    GERDING, DN
    PETERSON, LR
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1984, 20 (03) : 351 - 356