FLOW-INJECTION ANALYSIS WITH DETECTION AT MERCURY DROP ELECTRODE IN THE PRESENCE OF SURFACTANTS

被引:5
作者
KUBIAK, WW
KOWALSKI, Z
机构
[1] Faculty of Materials Science and Ceramics, Academy of Mining and Metallurgy, 30-059 Krakow
关键词
D O I
10.1016/0039-9140(94)E0020-R
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The surfactant presence in a sample may cause distortion of the flow injection peak obtained with amperometric detection at a mercury drop electrode. Distortion depends on detection parameters and flow system operating parameters and in some cases it may be eliminated by their careful optimization. Introduction of fumed silica into the carrier solution allows the dynamic removal of surfactant during the sample passage through the mixing coil. In the case of higher surfactant concentration in the sample, addition of fumed silica directly to the sample may be effective. Examples with amperometric detection as well as anodic stripping and adsorptive stripping voltammetric detection are described.
引用
收藏
页码:1319 / 1325
页数:7
相关论文
共 18 条
[1]  
BOND AM, 1984, ANAL CHIM ACTA, V162, P389, DOI 10.1016/S0003-2670(00)84263-4
[2]   FLOW-INJECTION CYCLIC VOLTAMMETRY [J].
CANETE, F ;
RIOS, A ;
DECASTRO, MDL ;
VALCARCEL, M .
ANALYTICA CHIMICA ACTA, 1988, 211 (1-2) :287-292
[3]   OZONE OXIDATION OF ORGANIC SEQUESTERING AGENTS IN WATER PRIOR TO DETERMINATION OF TRACE-METALS BY ANODIC-STRIPPING VOLTAMMETRY [J].
CLEM, RG ;
HODGSON, AT .
ANALYTICAL CHEMISTRY, 1978, 50 (01) :102-110
[4]   STRIPPING VOLTAMMETRY OF THIN MERCURY FILMS AT THE WALL-JET ELECTRODE [J].
GUNASINGHAM, H ;
ANG, KP ;
NGO, CC .
ANALYTICAL CHEMISTRY, 1985, 57 (02) :505-508
[5]  
KOWALSKI Z, 1984, EUROANALYSIS, V5
[6]   USE OF SILICA FOR ADSORPTIVE STRIPPING VOLTAMMETRY IN THE PRESENCE OF ORGANIC SURFACTANTS [J].
KUBIAK, W ;
WANG, J .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1989, 258 (01) :41-48
[7]   USE OF FUMED SILICA TO REMOVE SURFACTANT INTERFERENCES IN CONTINUOUS-FLOW POLAROGRAPHIC MEASUREMENTS [J].
KUBIAK, WW ;
KOWALSKI, Z .
ANALYTICAL CHEMISTRY, 1989, 61 (14) :1598-1600
[8]   ANODIC-STRIPPING VOLTAMMETRY OF HEAVY-METALS IN THE PRESENCE OF ORGANIC SURFACTANTS [J].
KUBIAK, WW ;
WANG, J .
TALANTA, 1989, 36 (08) :821-824
[9]   USE OF MERCURY DROP ELECTRODES FOR POLAROGRAPHIC MEASUREMENTS IN FLOWING SOLUTIONS [J].
KUBIAK, WW .
ELECTROANALYSIS, 1989, 1 (05) :379-388
[10]  
KUBIAK WW, UNPUB