摘要
The Collaborative Working Group (CWG) of the National Cancer Institute (NCI) electron beam treatment planning contract has performed a set of 14 experiments that measured dose distributions for 28 unique beam-phantom configurations that simulated various patient anatomic structures and beam geometries. Multiple dose distributions were measured with film or diode detectors for each configuration, resulting in 78, 2-D planar dose distributions and one, 1-D depth-dose distribution. Measurements were made for 9- and 20-MeV electron beams, using primarily 6 X 6- and 15 X 15-cm applicators at several SSDs. Dose distributions were measured for shaped fields, irregular surfaces, and inhomogeneities (1-D, 2-D, and 3-D), which were designed to simulate many clinical electron treatments. The data were corrected for asymmetries, and normalized in an absolute manner. This set of measured data can be used for verification of electron beam dose algorithms and is available to others for that purpose.
引用
收藏
页码:623 / 636
页数:14
相关论文
共 21 条
[1]  
BRAHME A, 1983, ACTA RADIOL S, V364, P101
[2]   A SOLID WATER PHANTOM MATERIAL FOR RADIOTHERAPY X-RAY AND GAMMA-RAY BEAM CALIBRATIONS [J].
CONSTANTINOU, C ;
ATTIX, FH ;
PALIWAL, BR .
MEDICAL PHYSICS, 1982, 9 (03) :436-441
[3]  
Constantinou C, 1978, THESIS U LONDON
[4]   A MODIFIED RADIOTHERAPY TECHNIQUE IN THE TREATMENT OF MEDULLOBLASTOMA [J].
DEWIT, L ;
VANDAM, J ;
RIJNDERS, A ;
VANDEVELDE, G ;
ANG, KK ;
VANDERSCHUEREN, E .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1984, 10 (02) :231-241
[5]   THE PROBLEM OF OBLIQUELY INCIDENT BEAMS IN ELECTRON-BEAM TREATMENT PLANNING [J].
EKSTRAND, KE ;
DIXON, RL .
MEDICAL PHYSICS, 1982, 9 (02) :276-278
[6]  
Hogstrom K R, 1983, Acta Radiol Suppl, V364, P89
[7]   ELECTRON-BEAM DOSE CALCULATIONS [J].
HOGSTROM, KR ;
MILLS, MD ;
ALMOND, PR .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (03) :445-459
[8]   DOSIMETRIC EVALUATION OF A PENCIL-BEAM ALGORITHM FOR ELECTRONS EMPLOYING A TWO-DIMENSIONAL HETEROGENEITY CORRECTION [J].
HOGSTROM, KR ;
MILLS, MD ;
MEYER, JA ;
PALTA, JR ;
MELLENBERG, DE ;
MEOZ, RT ;
FIELDS, RS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1984, 10 (04) :561-569
[9]  
HOGSTROM KR, 1983, MED PHYSICS MONOGRAP, V9, P223
[10]   RADIATION LEAKAGE FROM LINAC ELECTRON APPLICATOR ASSEMBLY [J].
KEYS, RA ;
PURDY, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1984, 10 (05) :713-721