SANDWICH-TYPE FLOW-THROUGH FIBEROPTIC CELLS FOR OPTICAL ABSORBENCY MEASUREMENTS

被引:12
作者
DASGUPTA, PK
BELLAMY, HS
LIU, HH
机构
[1] Department of Chemistry and Biochemistry, Texas Tech University, Lubbock
关键词
D O I
10.1016/0039-9140(93)80243-K
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The sandwich cell described by Pavon et al.1 and a similar sandwich cell, except with angled (45-degrees, confocal) single strand optical fibers and a conventional Z-type cell of 6-mm path length have been studied with respect to their performance for absorbance detection. Both sandwich cells show less susceptibility by one order of magnitude to artifact absorbance signals from RI changes than the Z-cell. The light throughput in the sandwich cells increase by an order of magnitude when an inert metallized reflector is used and this improves S/N. The overall light throughput is substantially greater for the angled entrance single strand fiber optic cell rather than the cell with the bifurcated fiber optic. Attainable limits of detection with these cells appear to be related to the pathlength for the cell dimensions studied.
引用
收藏
页码:341 / 345
页数:5
相关论文
共 7 条
  • [1] BETTERIDGE D, 1978, ANALYST, V103, P897, DOI 10.1039/an9780300897
  • [2] DASGUPTA PK, IN PRESS TALANTA
  • [3] DETERMINATION OF CREATININE IN UNDILUTED BLOOD-SERUM BY ENZYMATIC FLOW-INJECTION ANALYSIS WITH OPTOSENSING
    JEPPESEN, MT
    HANSEN, EH
    [J]. ANALYTICA CHIMICA ACTA, 1988, 214 (1-2) : 147 - 159
  • [4] PAVON JLP, 1992, ANAL CHEM, V64, P923
  • [5] RUZICKA J, 1985, ANAL CHIM ACTA, V173, P3
  • [6] OPTICS OF FLOW CELLS FOR LIQUID-CHROMATOGRAPHY
    STEWART, JE
    [J]. APPLIED OPTICS, 1981, 20 (04): : 654 - 659
  • [7] COMPENSATION OF THE SCHLIEREN EFFECT IN FLOW-INJECTION ANALYSIS BY USING DUAL-WAVELENGTH SPECTROPHOTOMETRY
    ZAGATTO, EAG
    ARRUDA, MAZ
    JACINTHO, AO
    MATTOS, IL
    [J]. ANALYTICA CHIMICA ACTA, 1990, 234 (01) : 153 - 160