LONG-TERM RELIABILITY OF SINGLE-LEAD ATRIAL SYNCHRONOUS PACING SYSTEMS USING CLOSELY SPACED ATRIAL DIPOLES - 5-YEAR EXPERIENCE

被引:31
作者
ANSANI, L
PERCOCO, GF
GUARDIGLI, G
TOSELLI, T
ANTONIOLI, GE
机构
[1] Divisione di Cardiologia Arcispedale, Ferrara, Anna
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1994年 / 17卷 / 11期
关键词
SINGLE LEAD VDD PACING; CLOSELY SPACED FLOATING ATRIAL DIPOLE;
D O I
10.1111/j.1540-8159.1994.tb03764.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To assess the long-term capability of single atrioventicular (AV) lead VDD pacing systems using close atrial dipoles to assure reliable atrial guided pacing, the safety and efficacy of 86 VDD units implanted in 73 patients at a single center since November 1988 was reviewed. All patients suffered from advanced AV block with normal sinoatrial function. Sixty five patients received a LEM/CCS Twinal 30/30S system, four patients received a Vitatron-Saphir system, and four patients received a Medtronic Thera VDR 8348 system. All patients underwent provocative tests in search of myopotential interference, and Holter recordings; in a group of patients who underwent pacemaker replacement a comparison was made between implant and replacement measurements. The mean follow-up duration was 27.3 months. A high percentage of successfully VDD paced patients and a low incidence of pacemaker malfunction, regularly solved by pacemaker reprogramming, was reported. Atrial signal amplitudes comparable to those measured at implant were found at replacement in all patients. These data support the long-term reliability of single AV lead VDD pacing systems with closely spaced atrial dipoles, as well as stable atrial sensing by floating bipolar atrial electrodes and effective atrial synchronous ventricular pacing over time.
引用
收藏
页码:1865 / 1869
页数:5
相关论文
共 14 条
[1]   ITALIAN MULTICENTER STUDY ON A SINGLE LEAD VDD PACING SYSTEM USING A NARROW ATRIAL DIPOLE SPACING [J].
ANTONIOLI, GE ;
ANSANI, L ;
BARBIERI, D ;
GUARDIGLI, G ;
PERCOCO, GF ;
TOSELLI, T .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11) :1890-1893
[2]  
ANTONIOLI GE, 1991, PACEMAKER LEADS 1991, P115
[3]  
ANTONIOLI GE, 1993, NEW PERSPECTIVES CAR, V3, P359
[4]  
ANTONIOLI GE, 1984, CARDIOSTIMULAZIONE, V2, P25
[5]  
ANTONIOLI GE, 1994, CARDIAC PACING ELECT, P209
[6]  
ANTONIOLI GE, 1979, CARDIAC PACING, pCH34
[7]  
BARBIERI D, 1991, PACEMAKER LEADS 1991, P135
[8]   EUROPEAN MULTICENTER PROSPECTIVE FOLLOW-UP-STUDY OF 1,002 IMPLANTS OF A SINGLE LEAD VDD PACING SYSTEM [J].
CRICK, JCP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :1742-1744
[9]   GUIDELINES FOR IMPLANTATION OF CARDIAC-PACEMAKERS AND ANTIARRHYTHMIA DEVICES - A REPORT OF THE AMERICAN-COLLEGE-OF-CARDIOLOGY AMERICAN-HEART-ASSOCIATION TASK-FORCE ON ASSESSMENT OF DIAGNOSTIC AND THERAPEUTIC CARDIOVASCULAR PROCEDURES (COMMITTEE ON PACEMAKER IMPLANTATION) [J].
DREIFUS, LS ;
FISCH, C ;
GRIFFIN, JC ;
GILLETTE, PC ;
MASON, JW ;
PARSONNET, V .
CIRCULATION, 1991, 84 (01) :455-467
[10]  
FURMAN S, 1991, PACEMAKER LEADS 1991, P183