ON THE CALIBRATION OF PLANE-PARALLEL IONIZATION CHAMBERS FOR ELECTRON-BEAM DOSIMETRY

被引:17
作者
ANDREO, P
RODRIGUES, LN
LINDBORG, L
KRAEPELIEN, T
机构
[1] SWEDISH RADIAT PROTECT INST,RADIAT STANDARDS LAB,S-10401 STOCKHOLM,SWEDEN
[2] KAROLINSKA HOSP,DEPT HOSP PHYS,S-10401 STOCKHOLM 60,SWEDEN
关键词
D O I
10.1088/0031-9155/37/5/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The procedure recommended by different dosimetry protocols for the determination of the absorbed dose to air chamber factor, N(D,pp), of plane-parallel chambers, comparing absorbed dose determinations in a high-energy electron beam with a reference cylindrical chamber having a known N(D,cyl) factor, has been investigated. Attention has been focused on the case that the chamber serving as reference has a solid aluminium central electrode. It has been found that using a wide spread Farmer-type chamber (NE 2571), together with recommendations which specifically take into account central electrode corrections for electron beam dosimetry, k(cel)p(cel) = p(cel-global)(IAEA) = 1.008, yields inconsistent results compared with those obtained from a fully homogeneous ionization chamber; for the N E 2571 chamber, a value k(cel)p(cel) = P(cel-global)(IAEA) congruent-to 1.0 has been obtained. Analytical calculations of k(m)k(att) for Farmer-type cylindrical chambers and experimental determinations of the product k(m)k(att)k(cel)p(cel) in electron beams agree within experimental uncertainties, with no evidence of statistical significance for the commonly used assumption p(cel) = 1, which yields a 0.8% correction (due to k(cel) only) for the effect of the NE 2571 aluminium electrode in electron beam dosimetry. The use of a 'NACP-chamber' specific factor (k(pp) or k(m)k(att)) to obtain N(D,pp) from N(K,pp) in NACP plane-parallel chambers has been found unsatisfactory, and direct experimental determinations of N(D,pp) are recommended instead. It is suggested that Standard Dosimetry Laboratories provide N(D,pp) calibration factors in Co-60 beams.
引用
收藏
页码:1147 / 1165
页数:19
相关论文
共 33 条
[1]   CHAMBER-DEPENDENT WALL CORRECTION FACTORS IN DOSIMETRY [J].
ANDREO, P ;
NAHUM, A ;
BRAHME, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (11) :1189-1199
[2]   ON THE CALIBRATION OF PLANE-PARALLEL ION CHAMBERS USING CO-60 BEAMS [J].
ANDREO, P ;
LINDBORG, L ;
MEDIN, J .
MEDICAL PHYSICS, 1991, 18 (02) :326-327
[3]   UNCERTAINTIES IN DOSIMETRIC DATA AND BEAM CALIBRATION [J].
ANDREO, P .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 19 (05) :1233-1247
[4]  
[Anonymous], 1983, MED PHYS, V10, P741
[6]  
BERGER MJ, 1964, NATL ACAD SCI NATL R, V1133, P205
[7]  
BOUTILLON M, 1973, METROLOGIA, V9, P139, DOI 10.1088/0026-1394/9/4/001
[8]  
GRINDBORG JE, 1991, SSI9104 SWED RAD PRO
[9]  
Johansson K.-A., 1978, National and International Standardization of Radiation Dosimetry, P243
[10]  
JOHANSSON KA, 1987, GURADFYS87 U GOT RAD, P13