A SIMPLE DEVICE FOR QUALITY-CONTROL OF POLYMER-FILMS, USING MEASUREMENTS OF PERMEATION AND DIFFUSION-COEFFICIENTS

被引:11
作者
MARCANDALLI, B [1 ]
SELLI, E [1 ]
TACCHI, R [1 ]
BELLOBONO, IR [1 ]
LEIDI, G [1 ]
机构
[1] BIAMATIC SRL,COMO,ITALY
关键词
DIFFUSION; -; Measurements; GASES; Diffusion; PERMEAMETERS; MECHANICAL PERMEABILITY - Applications - PLASTICS FILMS - Quality Control - POLYAMIDES - Permeability; Mechanical;
D O I
10.1016/0011-9164(84)85057-2
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
A simple device for the rapid determination of transport characteristics of vapors, gases or gaseous mixtures through polymer films was constructed. This device may be useful either for routine quality control or for precise research work. It consists of two coupled permeation cells: the difference between the permeation rates of a tested specimen with respect to a standard film is measured and the resulting signal interpreted in terms of diffusion and permeation coefficients. A theoretical analysis of the transient signal in terms of transport parameters is reported together with some experimental results on commercial polycaprolactam films.
引用
收藏
页码:113 / 122
页数:10
相关论文
共 6 条
[1]  
Barrer R., 1941, DIFFUSION SOLIDS, V1st
[2]  
Crank J, 1956, MATH DIFFUSION
[3]   ESTIMATION OF GAS TRANSPORT-COEFFICIENTS FROM DIFFERENTIAL PERMEATION, INTEGRAL PERMEATION, AND SORPTION RATE DATA [J].
FELDER, RM .
JOURNAL OF MEMBRANE SCIENCE, 1978, 3 (01) :15-27
[4]  
JAMES DI, 1969, BR POLYM J, V1, P205
[5]   A DYNAMIC APPROACH TO DIFFUSION AND PERMEATION MEASUREMENTS [J].
PASTERNAK, RA ;
SCHIMSCHEIMER, JF ;
HELLER, J .
JOURNAL OF POLYMER SCIENCE PART A-2-POLYMER PHYSICS, 1970, 8 (03) :467-+
[6]  
ZANDERIGHI L, 1979, GAZZ CHIM ITAL, V109, P505