MEASUREMENT OF COLUMN EFFICIENCY IN WHOLE COLUMN DETECTION CHROMATOGRAPHY

被引:13
作者
ROWLEN, KL
DUELL, KA
AVERY, JP
BIRKS, JW
机构
[1] UNIV COLORADO,DEPT CHEM,CAMPUS BOX 216,BOULDER,CO 80309
[2] UNIV COLORADO,COOPERAT INST RES ENVIRONM SCI,BOULDER,CO 80309
关键词
D O I
10.1021/ac00006a006
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The efficiency of a high-performance liquid chromatography (HPLC) column is measured independently of extra-column contributions to peak variance by means of whole column detection (WCD). The apparatus consists of a commercially available column instrumented with 14 individual UV absorbance detection zones. The variance of the peak at each detector is determined by fitting the peak to a gamma density function. A plot of total variance versus column position gives a straight line whose slope is equal to the height equivalent to a theoretical plate, H, and whose intercept is the sum of all contributions to extra-column variance. Variance measurements obtained on column for the three isomers of nitroaniline over the range 20-100% acetonitrile in water gave the dependences of H and extra-column variance on the capacity factor, k'. It was found that the extra-column variance is proportional to eta/(k' + 1)2, where eta is the viscosity of the mobile phase. This result is explained in terms of broadening within the connecting tubing by an amount proportional to the viscosity and focusing at the head of the column by 1/(k' + 1)2. Contributions to extra-column variance by the injector and electronic integration time are calculated from first principles and found to be small, while the average variance contribution from the detectors is deducted from the intercept of a plot of extra-column variance vs eta/(k' + 1)2.
引用
收藏
页码:575 / 579
页数:5
相关论文
共 11 条
[1]  
BIDLINGMEYER BA, 1984, ANAL CHEM, V56, P1583
[2]   ESTIMATION OF THE BAND BROADENING CONTRIBUTION OF HPLC EQUIPMENT TO COLUMN ELUTION PROFILES [J].
CLAESSENS, HA ;
CRAMERS, CA ;
KUYKEN, MAJ .
CHROMATOGRAPHIA, 1987, 23 (03) :189-194
[3]   EXCHANGE OF COMMENTS ON DATA ACQUISITION FOR CHROMATOGRAPHIC PEAKS [J].
DUELL, KA ;
AVERY, JP ;
ROWLEN, KL ;
BIRKS, JW .
ANALYTICAL CHEMISTRY, 1991, 63 (01) :73-75
[4]   DUAL ON-COLUMN FLUORESCENCE DETECTION SCHEME FOR CHARACTERIZATION OF CHROMATOGRAPHIC PEAKS [J].
EVANS, CE ;
MCGUFFIN, VL .
ANALYTICAL CHEMISTRY, 1988, 60 (06) :573-577
[5]  
GELDERLOOS DG, 1986, ANAL CHEM, V54, P900
[6]   SAMPLING AND EXTRA-COLUMN EFFECTS IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY - INFLUENCE OF PEAK SKEW ON PLATE COUNT CALCULATIONS [J].
KIRKLAND, JJ ;
YAU, WW ;
STOKLOSA, HJ ;
DILKS, CH .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1977, 15 (08) :303-316
[7]   WHOLE COLUMN DETECTION - APPLICATION TO HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
ROWLEN, KL ;
DUELL, KA ;
AVERY, JP ;
BIRKS, JW .
ANALYTICAL CHEMISTRY, 1989, 61 (23) :2624-2630
[8]   PHOTOCHEMICAL AMPLIFIER FOR LIQUID-CHROMATOGRAPHY BASED ON SINGLET OXYGEN SENSITIZATION [J].
SHELLUM, CL ;
BIRKS, JW .
ANALYTICAL CHEMISTRY, 1987, 59 (14) :1834-1841
[9]   THE ROLE OF THE THEORETICAL PLATE CONCEPT AND GAMMA DENSITY-FUNCTIONS IN STUDIES ON MASS-TRANSPORT AND MASS LOADABILITY OF CHROMATOGRAPHIC COLUMNS [J].
SMIT, HC ;
SMIT, JC ;
DEJAGER, EM .
CHROMATOGRAPHIA, 1986, 22 (1-6) :123-131
[10]  
Sternberg J., 1966, ADV CHROMATOGR, P205