BIOCOMPATIBILITY OF CHEMICAL-VAPOR-DEPOSITED DIAMOND

被引:244
作者
TANG, L
TSAI, C
GERBERICH, WW
KRUCKEBERG, L
KANIA, DR
机构
[1] UNIV MINNESOTA,BIOMED ENGN PROGRAM,MINNEAPOLIS,MN 55455
[2] UNIV MINNESOTA,DIGHT LABS,MINNEAPOLIS,MN 55455
[3] UNIV MINNESOTA,DEPT CHEM ENGN & MAT SCI,MINNEAPOLIS,MN 55455
[4] UNIV MINNESOTA,DEPT CHEM,MINNEAPOLIS,MN 55455
[5] CRYSTALLUME INC,MENLO PK,CA 94025
关键词
CHEMICAL VAPOR DEPOSITED DIAMOND; BIOCOMPATIBILITY; FIBRINOGEN; NEUTROPHILS; IMPLANTS;
D O I
10.1016/0142-9612(95)98822-V
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The biocompatibility of chemical-vapour-deposited (CVD) diamond surfaces has been assessed. Our results indicate that CVD diamond is as biocompatible as titanium (Ti) and 316 stainless steel (SS). First, the amount of adsorbed and 'denatured' fibrinogen on CVD diamond was very close to that of Ti and SS. Second, both in vitro and in vivo there appears to be less cellular adhesion and activation on the surface of CVD diamond surfaces compared to Ti and SS. This evident biocompatibility, coupled with the corrosion resistance and notable mechanical integrity of CVD diamond, suggests that diamond-coated surfaces may be highly desirable in a number of biomedical applications.