A 3-DIMENSIONAL FINITE-ELEMENT MODEL OF HUMAN TRANSTHORACIC DEFIBRILLATION - PADDLE PLACEMENT AND SIZE

被引:49
作者
CAMACHO, MA [1 ]
LEHR, JL [1 ]
EISENBERG, SR [1 ]
机构
[1] BRIGHAM & WOMENS HOSP,DEPT CLIN ENGN,BOSTON,MA 02115
基金
美国国家科学基金会;
关键词
D O I
10.1109/10.387196
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A detailed 3-D finite element model of the conductive anatomy of the human thorax has been constructed to quantitatively assess the current density distribution produced in the heart and thorax during transthoracic defibrillation. The model is based on a series of cross-sectional CT scans and incorporates isotropic conductivities for eight tissues and an approximation of the anisotropic conductivity of skeletal muscle, Current density distributions were determined and compared for four paddle pairs and two paddle sizes, Our results show that the myocardial current density distributions resulting from a defibrillation shock were fairly uniform for the paddle pairs and sizes examined in this study. Specific details of the spatial distribution of the current density magnitudes in the heart were found to depend on paddle placement and size, When the minimum current necessary to defibrillate was delivered, the maximum myocardial current density produced with any of the paddle sizes and positions examined was less than four times the minimum current density necessary to render a myocyte in a fibrillating heart inexcitable, and less than 40% of the damage threshold, These results suggest that common clinically used defibrillation paddle positions have a safety margin as large as 2.5 for current and similar to 6 for energy.
引用
收藏
页码:572 / 578
页数:7
相关论文
共 38 条
[1]   THERAPEUTIC INDEXES FOR TRANSCHEST DEFIBRILLATOR SHOCKS - EFFECTIVE, DAMAGING, AND LETHAL ELECTRICAL DOSES [J].
BABBS, CF ;
TACKER, WA ;
VANVLEET, JF ;
BOURLAND, JD ;
GEDDES, LA .
AMERICAN HEART JOURNAL, 1980, 99 (06) :734-738
[2]   APPLIED POTENTIAL TOMOGRAPHY [J].
BARBER, DC ;
BROWN, BH .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1984, 17 (09) :723-733
[3]   MECHANISM OF CARDIAC DEFIBRILLATION - A DIFFERENT POINT-OF-VIEW [J].
CHEN, PS ;
WOLF, PD ;
IDEKER, RE .
CIRCULATION, 1991, 84 (02) :913-919
[4]   A VOLUME CONDUCTOR MODEL OF THE THORAX FOR THE STUDY OF DEFIBRILLATION FIELDS [J].
CLAYDON, FJ ;
PILKINGTON, TC ;
TANG, ASL ;
MORROW, MN ;
IDEKER, RE .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (11) :981-992
[5]   MYOCARDIAL NECROSIS FROM DIRECT-CURRENT COUNTERSHOCK - EFFECT OF PADDLE ELECTRODE SIZE AND TIME INTERVAL BETWEEN DISCHARGES [J].
DAHL, CF ;
EWY, GA ;
WARNER, ED ;
THOMAS, ED .
CIRCULATION, 1974, 50 (05) :956-961
[6]   ELECTRODE PAD SIZE, TRANS-THORACIC IMPEDANCE AND SUCCESS OF EXTERNAL VENTRICULAR DEFIBRILLATION [J].
DALZELL, GWN ;
CUNNINGHAM, SR ;
ANDERSON, J ;
ADGEY, AAJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 64 (12) :741-744
[7]  
DEJONGH AL, 1993, P IEEE ENG MED BIOL, V15, P828
[8]   CARDIAC DAMAGE PRODUCED BY DIRECT-CURRENT COUNTER-SHOCK APPLIED TO THE HEART [J].
DOHERTY, PW ;
MCLAUGHLIN, PR ;
BILLINGHAM, M ;
KERNOFF, R ;
GORIS, ML ;
HARRISON, DC .
AMERICAN JOURNAL OF CARDIOLOGY, 1979, 43 (02) :225-232
[9]  
FAHY JB, 1987, IEEE T BIOMED ENG, V34, P743
[10]   FUNDAMENTAL CRITERIA UNDERLYING THE EFFICACY AND SAFETY OF DEFIBRILLATING CURRENT WAVEFORMS [J].
GEDDES, LA ;
NIEBAUER, MJ ;
BABBS, CF ;
BOURLAND, JD .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1985, 23 (02) :122-130