FAST EVALUATION OF PATIENT SET-UP DURING RADIOTHERAPY BY ALIGNING FEATURES IN PORTAL AND SIMULATOR IMAGES

被引:70
作者
BIJHOLD, J
VANHERK, M
VIJLBRIEF, R
LEBESQUE, JV
机构
[1] Netherlands Cancer Inst., Amsterdam
关键词
D O I
10.1088/0031-9155/36/12/010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new fast method is presented for the quantification of patient set-up errors during radiotherapy with external photon beams. The set-up errors are described as deviations in relative position and orientation of specified anatomical structures relative to specified field shaping devices. These deviations are determined from parameters of the image transformations that make their features in a portal image align with the corresponding features in a simulator image. Knowledge of some set-up parameters during treatment simulation is required. The method does not require accurate knowledge about the position of the portal imaging device as long as the positions of some of the field shaping devices are verified independently during treatment. By applying this method, deviations in a pelvic phantom set-up can be measured with a precision of 2 mm within 1 minute. Theoretical considerations and experiments have shown that the method is not applicable when there are out-of-plane rotations larger than 2 degrees or translations larger than 1 cm. Interobserver variability proved to be a source of large systematic errors, which could be reduced by offering a precise protocol for the feature alignment.
引用
收藏
页码:1665 / 1679
页数:15
相关论文
共 9 条
  • [1] RADIATION-FIELD EDGE-DETECTION IN PORTAL IMAGES
    BIJHOLD, J
    GILHUIJS, KGA
    VANHERK, M
    MEERTENS, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (12) : 1705 - 1710
  • [2] ACCURACY IN RADIATION-FIELD ALIGNMENT IN HEAD AND NECK-CANCER - A PROSPECTIVE-STUDY
    HUIZENGA, H
    LEVENDAG, PC
    DEPORRE, PMZR
    VISSER, AG
    [J]. RADIOTHERAPY AND ONCOLOGY, 1988, 11 (02) : 181 - 187
  • [3] 1ST CLINICAL-EXPERIENCE WITH A NEWLY DEVELOPED ELECTRONIC PORTAL IMAGING DEVICE
    MEERTENS, H
    VANHERK, M
    BIJHOLD, J
    BARTELINK, H
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (05): : 1173 - 1181
  • [4] A METHOD FOR THE MEASUREMENT OF FIELD PLACEMENT ERRORS IN DIGITAL PORTAL IMAGES
    MEERTENS, H
    BIJHOLD, J
    STRACKEE, J
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (03) : 299 - 323
  • [5] ACCURACY OF RADIATION-FIELD ALIGNMENT IN CLINICAL-PRACTICE
    RABINOWITZ, I
    BROOMBERG, J
    GOITEIN, M
    MCCARTHY, K
    LEONG, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (10): : 1857 - 1867
  • [6] SOLUTION TO TREATMENT PLANNING PROBLEMS USING COORDINATE TRANSFORMATIONS
    SIDDON, RL
    [J]. MEDICAL PHYSICS, 1981, 8 (06) : 766 - 774
  • [7] A MATRIX IONIZATION-CHAMBER IMAGING DEVICE FOR ONLINE PATIENT SETUP VERIFICATION DURING RADIOTHERAPY
    VANHERK, M
    MEERTENS, H
    [J]. RADIOTHERAPY AND ONCOLOGY, 1988, 11 (04) : 369 - 378
  • [8] VANHERK M, 1990, 10TH P INT C US COMP, P95
  • [9] 1976, ICRU24 REP