INSTRUMENTAL CONDITIONS OF SECONDARY ION MASS-SPECTROMETRY THAT AFFECT SENSITIVITY FOR OBSERVATION OF VERY HIGH MASSES

被引:23
作者
ABERTH, W
机构
[1] Univ of California, San Francisco,, CA, USA, Univ of California, San Francisco, CA, USA
关键词
BIOLOGICAL MATERIALS - Spectroscopic Analysis - SPECTROSCOPY - Applications;
D O I
10.1021/ac00297a055
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
It is demonstrated by comparing the mass spectra of cesium iodide cluster ions obtained under different operating parameters that such factors as acceleration voltage, system pressure, and primary beam energy affect the high-mass signal current to a greater extent than that of the low mass. The measurements imply that consideration should be given to these parameters for more effective analysis of high-mass compounds.