THEORETICAL AND EXPERIMENTAL COMPARISON OF 3 TYPES OF ELECTROMAGNETIC HYPERTHERMIA APPLICATOR

被引:7
作者
JOHNSON, RH
PREECE, AW
GREEN, JL
机构
[1] Oncology Res. Univ., Bristol R. Infirmary
关键词
D O I
10.1088/0031-9155/35/6/006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There is evidence that heating of malignant tissue in the treatment of cancer may be beneficial and so the performance of different applicator designs needs to be established. A theoretical model of the dielectric loaded waveguide applicator is compared with models of two other applicators which depend on energy radiated from conductors carrying high frequency current. The latter are exemplified by the compact resonant patch applicator and the lightweight inductive current sheet applicator. It is shown that heating profiles and field penetration of each applicator are similar for equal radiating areas, and these results have been substantiated experimentally. Impedance match as a function of frequency and load is also compared for the three types of applicator.
引用
收藏
页码:761 / 779
页数:19
相关论文
共 10 条
[1]  
CHOU CK, 1984, BIOELECTROMAGNETICS, V5, P435, DOI 10.1002/bem.2250050408
[2]  
EDWARDS TC, 1981, F MICROSTRIP CIRCUIT, P45
[3]  
HAND J, 1986, PHYSICAL TECHNIQUES
[4]   NEW LOW-PROFILE APPLICATORS FOR LOCAL HEATING OF TISSUES [J].
JOHNSON, RH ;
JAMES, JR ;
HAND, JW ;
HOPEWELL, JW ;
DUNLOP, PRC ;
DICKINSON, RJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (01) :28-37
[6]  
JOHNSON RH, 1988, IEEE MTT, V35, P1317
[7]  
KANTOR G, 1981, J MICROWAVE POWER EE, V16, P135
[8]  
PAGLIONE R, 1981, MICROWAVE J, V24, P71
[9]   THE USE OF AN INFRARED CAMERA FOR IMAGING THE HEATING EFFECT OF RF APPLICATORS [J].
PREECE, AW ;
MURFIN, JL .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1987, 3 (02) :119-122
[10]   EQUIVALENT CAPACITANCES OF MICROSTRIP OPEN CIRCUITS [J].
SILVESTER, P ;
BENEDEK, P .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1972, MT20 (08) :511-+