THE ACCURACY OF FINAPRES NONINVASIVE MEAN ARTERIAL-PRESSURE MEASUREMENTS IN ANESTHETIZED PATIENTS

被引:56
作者
GIBBS, NM
LARACH, DR
DERR, JA
机构
[1] PENN STATE UNIV, COLL SCI, CTR BIOSTAT, UNIVERSITY PK, PA 16802 USA
[2] PENN STATE UNIV, COLL SCI, DEPT STAT, UNIVERSITY PK, PA 16802 USA
关键词
MEASUREMENT TECHNIQUE; BLOOD PRESSURE; NONINVASIVE EQUIPMENT; ACCURACY; BIAS; FINAPRES;
D O I
10.1097/00000542-199104000-00004
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The Finapres(TM) (FIN) is a new noninvasive blood pressure monitor that provides continuous arterial waveform display with the use of a finger cuff. The authors assessed the accuracy of FIN mean arterial pressure (MAP) measurements relative to simultaneous direct radial arterial pressures in 20 patients undergoing general anesthesia for major elective surgery. Data were collected digitally with the use of RS-232 communications over a total of 16.2 h. The data were processed into 6012 interference-free time samples, each spanning 6 s. The authors determined the difference between FIN and direct MAPs during each time sample. The authors calculated not only the bias of FIN measurements, but also the frequency, magnitude, and duration of discrepancies between simultaneous FIN and direct MAPs. The overall bias of the FIN MAP was -0.5 +/- 1.0 mmHg, which was not significantly different from zero. However, 32.3 +/- 6.2% of all MAP comparisons differed by greater than +/- 10 mmHg, and 5.0 +/- 1.1% differed by greater than +/- 20 mmHg. Moreover, there was an average of one episode every 2 patient-hours when the FIN MAP differed by greater than +/- 20 mmHg for more than 1 min. Although the MAP measured by FIN accurately reflected direct MAPs most of the time, there were occasional discrepancies of different magnitude such that clinical usefulness may be limited in patients in whom continuous accurate blood pressure measurements are essential.
引用
收藏
页码:647 / 652
页数:6
相关论文
共 21 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] BOEHMER RD, 1987, J CLIN MONITOR, V3, P282
  • [3] BRUNER JMR, 1978, HDB BLOOD PRESSURE M, P94
  • [4] PLASMA BUPIVACAINE CONCENTRATIONS FOLLOWING ILIOINGUINAL-ILIOHYPOGASTRIC NERVE BLOCKADE IN CHILDREN
    EPSTEIN, RH
    LARIJANI, GE
    WOLFSON, PJ
    ALAKOKKO, TI
    BOERNER, TF
    [J]. ANESTHESIOLOGY, 1988, 69 (05) : 773 - 776
  • [5] DIRECT BLOOD-PRESSURE MEASUREMENT - DYNAMIC-RESPONSE REQUIREMENTS
    GARDNER, RM
    [J]. ANESTHESIOLOGY, 1981, 54 (03) : 227 - 236
  • [6] GORBACK M S, 1988, Anesthesiology (Hagerstown), V69, pA325, DOI 10.1097/00000542-198809010-00324
  • [7] GUYTON A. C., 1986, TXB MED PHYSL, P244
  • [8] HUFFNAGLE S, 1989, Anesthesiology (Hagerstown), V71, pA405, DOI 10.1097/00000542-198909001-00405
  • [9] HUG CC, 1986, ANESTHESIA, P411
  • [10] COMPARISON OF THE FINAPRES BLOOD-PRESSURE MONITOR WITH INTRA-ARTERIAL MANOMETRY DURING INDUCTION OF ANESTHESIA
    KERMODE, JL
    DAVIS, NJ
    THOMPSON, WR
    [J]. ANAESTHESIA AND INTENSIVE CARE, 1989, 17 (04) : 470 - 475