A GOLD PLATINUM THERMOCOUPLE REFERENCE TABLE

被引:21
作者
GOTOH, M [1 ]
HILL, KD [1 ]
MURDOCK, EG [1 ]
机构
[1] TAMAGAWA UNIV,TOKYO 194,JAPAN
关键词
D O I
10.1063/1.1142213
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The thermoelectric voltage of the gold/platinum thermocouple has been measured over the range 0-962-degrees-C by comparison with calibrated platinum resistance thermometers. From 0 to 500-degrees-C, the measurements were carried out in stirred liquid baths; from 660 to 964-degrees-C, they were carried out in a pressure-controlled sodium-filled heat pipe furnace that provided an isothermal intercomparison environment. Measurements were also made in the metal freezing points of indium, tin, zinc, antimony, and silver, at the melting point of gallium, and at the liquidus point of the silver-copper eutectic. By fitting the measured thermoelectric voltages to a single eighth-degree polynomial in temperature by the method of least squares, a reference function is obtained for the Au/Pt thermocouple that provides emf as a function of temperature (ITS-90) to within +/- 10 mK from 0 to 962-degrees-C. The Au/Pt thermocouple merits serious consideration for precise temperature measurements as its stability approaches that of the high-temperature platinum resistance thermometer.
引用
收藏
页码:2778 / 2791
页数:14
相关论文
共 26 条
[1]  
BASSANI C, 1980, HIGH TEMP HIGH PRESS, V12, P351
[2]   RECOMMENDED VALUES OF TEMPERATURE FOR A SELECTED SET OF SECONDARY REFERENCE POINTS [J].
BEDFORD, RE ;
BONNIER, G ;
MAAS, H ;
PAVESE, F .
METROLOGIA, 1984, 20 (04) :145-155
[3]  
BERRY RJ, 1988, NRCC29860 NAT RES CO
[4]  
BERRY RJ, 1988, NRCC29861 NAT RES CO
[5]  
BERRY RJ, 1989, NRCC2986 NAT RES COU
[6]   COMPARISON OF RESISTANCE VERSUS THERMODYNAMIC TEMPERATURE OF PLATINUM RESISTANCE THERMOMETERS WITH THE ITS-90 [J].
CONNOLLY, JJ ;
JONES, TP ;
TAPPING, J .
METROLOGIA, 1990, 27 (02) :83-88
[7]   THE ELECTRICAL-CONDUCTIVITY OF WEIGHT DILUTED AND CONCENTRATED STANDARD SEAWATER AS A FUNCTION OF SALINITY AND TEMPERATURE [J].
DAUPHINEE, TM ;
ANCSIN, J ;
KLEIN, HP ;
PHILLIPS, MJ .
IEEE JOURNAL OF OCEANIC ENGINEERING, 1980, 5 (01) :28-41
[8]  
HILL KD, 1989, MINIMIZATION SUBRANG
[9]   AC-INSULATION AND DC-INSULATION LEAKAGE IN PLATINUM RESISTANCE THERMOMETERS UP TO 750-DEGREES-C [J].
JIPEI, Z ;
BERRY, RJ .
METROLOGIA, 1985, 21 (04) :207-223
[10]  
KIRBY CGM, 1982, TEMP ITS MEAS, V5, P1293