DEVELOPMENT OF A NEEDLE-TYPE BIOSENSOR FOR INTRAVASCULAR GLUCOSE MONITORING

被引:24
作者
ABDELHAMID, I [1 ]
ATANASOV, P [1 ]
WILKINS, E [1 ]
机构
[1] UNIV NEW MEXICO,DEPT CHEM & NUCL ENGN,ALBUQUERQUE,NM 87131
关键词
BIOSENSORS; GLUCOSE; PHENYLENEDIAMINE; POLYMER COATINGS;
D O I
10.1016/0003-2670(95)00172-V
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A needle type glucose biosensor has been developed for monitoring glucose levels during the hemorrhagic shock in trauma patients. The sensor employs polyurethane, cellulose acetate or PVC layers as an outer glucose diffusion and protective membrane, and glucose oxidase enzyme entrapped in poly(1,3-phenylenediamine) film. Sensor performance was evaluated in vitro and the sensor shows a sensitivity of up to 35 nA/mM and a linear range of up to 38 mM. Evaluation of the sensor response in serum showed similar sensitivity and linear range as obtained using calibration curves in buffer solution. The sensor has a short response time of 23 s. The sensors were operated continuously for 14 days in phosphate buffer solution, and no significant change in the sensitivity and the linear range was observed during the first 5 days. Sensors show a minimum change in their performance when stored inactive in buffer at 4 degrees C for at least eight weeks.
引用
收藏
页码:45 / 54
页数:10
相关论文
共 29 条
[1]   ELECTROCHEMICAL IMMOBILIZATION OF ENZYMES .5. MICROELECTRODES FOR THE DETECTION OF GLUCOSE BASED ON GLUCOSE-OXIDASE IMMOBILIZED IN A POLY(PHENOL) FILM [J].
BARTLETT, PN ;
CARUANA, DJ .
ANALYST, 1992, 117 (08) :1287-1292
[2]  
BARTLETT PN, 1993, J ELECTROANAL CHEM, V362, P1
[3]   FLOW-INJECTION ANALYSIS OF GLUCOSE AT AN AMPEROMETRIC GLUCOSE SENSOR-BASED ON ELECTROCHEMICAL CODEPOSITION OF PALLADIUM AND GLUCOSE-OXIDASE ON A GLASSY-CARBON ELECTRODE [J].
CHI, QJ ;
DONG, SJ .
ANALYTICA CHIMICA ACTA, 1993, 278 (01) :17-23
[4]   ENZYME ENTRAPMENT IN ELECTRICALLY CONDUCTING POLYMERS - IMMOBILIZATION OF GLUCOSE-OXIDASE IN POLYPYRROLE AND ITS APPLICATION IN AMPEROMETRIC GLUCOSE SENSORS [J].
FOULDS, NC ;
LOWE, CR .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS I, 1986, 82 :1259-1264
[5]   IMMOBILIZATION OF GLUCOSE-OXIDASE IN FERROCENE-MODIFIED PYRROLE POLYMERS [J].
FOULDS, NC ;
LOWE, CR .
ANALYTICAL CHEMISTRY, 1988, 60 (22) :2473-2478
[6]   ELECTROPOLYMERIZED 1,3-DIAMINOBENZENE FOR THE CONSTRUCTION OF A 1,1'-DIMETHYLFERROCENE MEDIATED GLUCOSE BIOSENSOR [J].
GEISE, RJ ;
RAO, SY ;
YACYNYCH, AM .
ANALYTICA CHIMICA ACTA, 1993, 281 (03) :467-473
[7]   POLYMER FILM CHEMICALLY MODIFIED ELECTRODE AS A POTENTIOMETRIC SENSOR [J].
HEINEMAN, WR ;
WIECK, HJ ;
YACYNYCH, AM .
ANALYTICAL CHEMISTRY, 1980, 52 (02) :345-346
[8]  
IKARIYAMA Y, 1987, ANAL LETT, V20, P1791
[9]   GLUCOSE SENSITIVITY OF POLYPYRROLE FILMS CONTAINING IMMOBILIZED GLUCOSE-OXIDASE AND HYDROQUINONESULFONATE IONS [J].
KAJIYA, Y ;
SUGAI, H ;
IWAKURA, C ;
YONEYAMA, H .
ANALYTICAL CHEMISTRY, 1991, 63 (01) :49-54
[10]   GLUCOSE FAST-RESPONSE AMPEROMETRIC SENSOR BASED ON GLUCOSE-OXIDASE IMMOBILIZED IN AN ELECTROPOLYMERIZED POLY(ORTHO-PHENYLENEDIAMINE) FILM [J].
MALITESTA, C ;
PALMISANO, F ;
TORSI, L ;
ZAMBONIN, PG .
ANALYTICAL CHEMISTRY, 1990, 62 (24) :2735-2740