PULSATILE FLOW MODEL WITH ELASTIC VESSELS FOR DUPLEX SONOGRAPHIC EXAMINATIONS

被引:5
作者
PETRICK, J [1 ]
SCHLIEF, R [1 ]
ZOMACK, M [1 ]
LANGHOLZ, J [1 ]
URBANK, A [1 ]
机构
[1] FRANZISKUS KRANKENHAUS BERLIN, BERLIN, GERMANY
来源
ULTRASCHALL IN DER MEDIZIN | 1992年 / 13卷 / 06期
关键词
FLOW MODEL; PULSATILE FLOW; DUPLEX SONOGRAPHY; DOPPLER SPECTRUM;
D O I
10.1055/s-2007-1005322
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Using ultrasound duplex technique flow phenomena in patients' circulation can be examined. For the interpretation of these examinations it is necessary to have extensive knowledge on flow influencing parameters. This can be easily obtained from simplified flow models. This article describes the components of a flow model that allows examination of ultrasonic contrast media flowing through an artificial heart and vessel mimicking tubes. The artificial heart is the drive which pumps a water glycerol cellulose mixture through the circulation in a pulsatile manner. The shape of the ventricle, the compliance of the aorta, the viscosity of the flow medium and the wall elasticity of the examination vessel were taken into account. The attenuation caused by the surrounding tissue is simulated by a variable layer of castor oil. The flow model is suitable to produce flow profiles that are very similar to physiological profiles.
引用
收藏
页码:277 / 282
页数:6
相关论文
共 20 条
[1]   SIMULATION OF ARTERIAL PULSATILE FLOW IN ELASTIC TUBES AND IN A CAROTID-ARTERY SPECIMEN [J].
ARNOLDS, BJ ;
KAPP, H ;
VONREUTERN, GM .
ULTRASCHALL IN DER MEDIZIN, 1984, 5 (05) :231-236
[2]   IMPROVED SENSITIVITY OF COLOR DOPPLER BY SH-U-454 [J].
BECHER, H ;
SCHLIEF, R .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 64 (05) :374-377
[3]   PERFORMANCE TESTS OF DOPPLER ULTRASOUND EQUIPMENT WITH A TISSUE AND BLOOD-MIMICKING PHANTOM [J].
BOOTE, EJ ;
ZAGZEBSKI, JA .
JOURNAL OF ULTRASOUND IN MEDICINE, 1988, 7 (03) :137-147
[4]   AN INVIVO-LIKE PHANTOM FOR EVALUATING THE CLINICAL USE OF DUPLEX SCANNING AND COLOR-CODED DOPPLER DEVICES (CCDD) [J].
BRASSEL, F ;
HENDRICKX, P ;
ROTH, U ;
FROHLICH, H ;
WAGNER, HH .
ULTRASCHALL IN DER MEDIZIN, 1990, 11 (06) :290-294
[5]   AN INVITRO MODEL AND ITS APPLICATION FOR THE STUDY OF CAROTID DOPPLER SPECTRAL BROADENING [J].
DOUVILLE, Y ;
JOHNSTON, KW ;
KASSAM, M ;
ZUECH, P ;
COBBOLD, RSC ;
JARES, A .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1983, 9 (04) :347-356
[6]  
Evans D. H., 1989, DOPPLER ULTRASOUND P
[7]  
Hatle L, 1985, DOPPLER ULTRASOUND C, V2nd
[8]  
HEISS HW, 1990, CORVAS, V3, P127
[9]   A COMPARISON OF THE DOPPLER SPECTRA FROM HUMAN BLOOD AND ARTIFICIAL BLOOD USED IN A FLOW PHANTOM [J].
HOSKINS, PR ;
LOUPAS, T ;
MCDICKEN, WN .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (02) :141-147
[10]  
KNOTT E, 1988, J THORAC CARDIOV SUR, V96, P952