MAGNETIC-RESONANCE-IMAGING ANALYSIS OF EXTREMELY SLOW FLOW IN A MODEL SHUNT SYSTEM

被引:13
作者
FRANK, E
BUONOCORE, M
HEIN, L
机构
[1] UNIV CALIF DAVIS,SACRAMENTO MED CTR,DEPT RADIOL,SACRAMENTO,CA 95817
[2] UNIV CALIF DAVIS,SACRAMENTO MED CTR,DEPT NEUROSURG,SACRAMENTO,CA 95817
关键词
MAGNETIC RESONANCE IMAGING; SHUNT MALFUNCTION; CEREBROSPINAL FLUID SHUNT; CEREBROSPINAL FLUID;
D O I
10.1007/BF00298443
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Shunt malfunction is common and its diagnosis may require invasive testing that may be inaccurate or result in complications. Magnetic resonance imaging (MRI) may prove to be a useful noninvasive test of shunt function as it has been shown that MRI is capable of measuring cerebrospinal fluid (CSF) flows from 2 ml/h to 40 ml/h in model systems. Since flows in functioning shunt systems can be less than 2 ml/h, MRI must be sensitive enough to detect flow in this range in order to be a valid test for shunt function. Continuing previous studies, we have studied MRI flow-related enhancement at flow rates from 0 to 2 ml/h. Multiple spin echo scans (TR2000, TE20) were made through a specialized section of tubing in a model shunt system. The intensity of the MRI signal at points known to demonstrate maximal flow-related enhancement was measured. A linear relationship was demonstrated between signal intensity and flow as low as 0.8 ml/h. These results add support to the concept that MRI is sensitive enough to detect the lowest flows present in functioning shunt systems and therefore may be useful as a noninvasive test of shunt function.
引用
收藏
页码:73 / 75
页数:3
相关论文
共 13 条
[1]   THE APPEARANCE OF RAPIDLY FLOWING BLOOD ON MAGNETIC-RESONANCE IMAGES [J].
BRADLEY, WG ;
WALUCH, V ;
LAI, KS ;
FERNANDEZ, EJ ;
SPALTER, C .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 143 (06) :1167-1174
[2]   FLOWING CEREBROSPINAL-FLUID IN NORMAL AND HYDROCEPHALIC STATES - APPEARANCE ON MR IMAGES [J].
BRADLEY, WG ;
KORTMAN, KE ;
BURGOYNE, B .
RADIOLOGY, 1986, 159 (03) :611-616
[3]   FLOW PHENOMENA IN MR IMAGING [J].
BRADLEY, WG .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 150 (05) :983-994
[4]  
CITRIN CM, 1986, AM J NEURORADIOL, V7, P1021
[5]  
FRANK E, 1990, British Journal of Neurosurgery, V4, P53, DOI 10.3109/02688699009000682
[6]   THE USE OF SHUNTING DEVICES FOR CEREBROSPINAL-FLUID IN CANADA [J].
HOFFMAN, HJ ;
SMITH, MSM .
CANADIAN JOURNAL OF NEUROLOGICAL SCIENCES, 1986, 13 (02) :81-87
[7]   A RADIONUCLIDE METHOD OF EVALUATING SHUNT FUNCTION AND CSF CIRCULATION IN HYDROCEPHALUS [J].
HOWMANGILES, R ;
MCLAUGHLIN, A ;
JOHNSTON, I ;
WHITTLE, I .
JOURNAL OF NEUROSURGERY, 1984, 61 (03) :604-605
[8]   CEREBROSPINAL-FLUID SHUNTS - FLOW MEASUREMENTS WITH MR IMAGING [J].
MARTIN, AJ ;
DRAKE, JM ;
LEMAIRE, C ;
HENKELMAN, RM .
RADIOLOGY, 1989, 173 (01) :243-247
[9]   DYNAMIC CHANGES OF CEREBROSPINAL-FLUID SHUNT FLOW IN PATIENTS DAILY LIFE [J].
MATSUMAE, M ;
MURAKAMI, T ;
UEDA, M ;
SUZUKI, Y ;
SATO, O .
CHILDS NERVOUS SYSTEM, 1987, 3 (01) :30-34
[10]   NUCLEAR MAGNETIC-RESONANCE - PRINCIPLES OF BLOOD-FLOW IMAGING [J].
MILLS, CM ;
BRANTZAWADZKI, M ;
CROOKS, LE ;
KAUFMAN, L ;
SHELDON, P ;
NORMAN, D ;
BANK, W ;
NEWTON, TH .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 142 (01) :165-170