QUANTITATIVE COMPUTED TOMOGRAPHY;
BONE MINERALS;
TECHNIQUE;
DOSIMETRY;
CLINICAL APPLICATIONS;
D O I:
10.1259/0007-1285-64-760-298
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 [临床医学];
100207 [影像医学与核医学];
1009 [特种医学];
摘要:
A single KV(p) quantitative CT (QCT) technique was used for measuring the spinal trabecular bone density. The trabecular bone equivalent density (TBED) was expressed as the density of a K2HPO4 solution that exhibits a linear attenuation coefficient identical to that of trabecular bone. A field non-uniformity correcting factor was obtained from experiments on phantoms. The effective equivalent dose of the whole examination (four sections and a scout view) is 370-mu-Sv. The in vivo short-term precision (reproducibility coefficient of variation) ranges from 1.4% to 4.1% depending on the TBED values of the normal subjects. The TBED was measured in 206 normal Greeks aged 30-69 years. Average TBED decreased with increasing age for both sexes. Analysis considering separately each vertebra showed a tendency to a caudal spinal TBED reduction. No correlation was found between the TBED values and the body habitus, milk consumption, smoking habits in men, and number of full-term pregnancies. A significant difference (p < 0.01) was found between the TBED values of the normals and those of 50 women suffering from post-menopausal osteoporosis, 37 alcoholic men, and 12 gastrectomized men (Billroth II). Quantitative CT has been established as a method of measuring TBED in health and disease, and the results from this study confirm these applications.