VISCOELASTICITY OF LIVING CELLS ALLOWS HIGH-RESOLUTION IMAGING BY TAPPING MODE ATOMIC-FORCE MICROSCOPY

被引:171
作者
PUTMAN, CAJ
VANDERWERF, KO
DEGROOTH, BG
VANHULST, NF
GREVE, J
机构
[1] Department of Applied Physics, University of Twente, Enschede
关键词
D O I
10.1016/S0006-3495(94)80649-6
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Application of atomic force microscopy (AFM) to biological objects and processes under physiological conditions has been hampered so far by the deformation and destruction of the soft biological materials invoked. Here we describe a new mode of operation in which the standard V-shaped silicon nitride cantilever is oscillated under liquid and damped by the interaction between AFM tip and sample surface. Because of the viscoelastic behavior of the cellular surface, cells effectively ''harden'' under such a tapping motion at high frequencies and become less susceptible to deformation. Images obtained in this way primarily reveal the surface structure of the cell. It is now possible to study physiological processes, such as cell growth, with a minimal level of perturbation and high spatial resolution (similar to 20 nm).
引用
收藏
页码:1749 / 1753
页数:5
相关论文
共 18 条
[1]  
AMER NM, 1988, APPL PHYS LETT, V53, P2400
[2]   ATOMIC FORCE MICROSCOPE [J].
BINNIG, G ;
QUATE, CF ;
GERBER, C .
PHYSICAL REVIEW LETTERS, 1986, 56 (09) :930-933
[3]   IMAGING CELLS WITH THE ATOMIC FORCE MICROSCOPE [J].
BUTT, HJ ;
WOLFF, EK ;
GOULD, SAC ;
NORTHERN, BD ;
PETERSON, CM ;
HANSMA, PK .
JOURNAL OF STRUCTURAL BIOLOGY, 1990, 105 (1-3) :54-61
[4]   INVITRO ACTIVATION OF HUMAN PLATELETS TRIGGERED AND PROBED BY ATOMIC FORCE MICROSCOPY [J].
FRITZ, M ;
RADMACHER, M ;
GAUB, HE .
EXPERIMENTAL CELL RESEARCH, 1993, 205 (01) :187-190
[5]   INSITU INVESTIGATIONS OF SINGLE LIVING CELLS INFECTED BY VIRUSES [J].
HABERLE, W ;
HORBER, JKH ;
OHNESORGE, F ;
SMITH, DPE ;
BINNIG, G .
ULTRAMICROSCOPY, 1992, 42 :1161-1167
[6]   ACTIN FILAMENT DYNAMICS IN LIVING GLIAL-CELLS IMAGED BY ATOMIC FORCE MICROSCOPY [J].
HENDERSON, E ;
HAYDON, PG ;
SAKAGUCHI, DS .
SCIENCE, 1992, 257 (5078) :1944-1946
[7]   VISCOSITY OF PASSIVE HUMAN NEUTROPHILS UNDERGOING SMALL DEFORMATIONS [J].
HOCHMUTH, RM ;
TINGBEALL, HP ;
BEATY, BB ;
NEEDHAM, D ;
TRANSONTAY, R .
BIOPHYSICAL JOURNAL, 1993, 64 (05) :1596-1601
[8]  
PARPURA V, 1993, J CELL SCI, V104, P427
[9]   ATOMIC FORCE MICROSCOPY OF BIOLOGICAL SAMPLES AT LOW-TEMPERATURE [J].
PRATER, CB ;
WILSON, MR ;
GARNAES, J ;
MASSIE, J ;
ELINGS, VB ;
HANSMA, PK .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1991, 9 (02) :989-991
[10]  
Putman C. A. J., 1993, Bioimaging, V1, P63, DOI 10.1002/1361-6374(199306)1:2<63::AID-BIO1>3.3.CO