EXTRUSION OF EYE SOCKET MAGNETIC IMPLANT AFTER MR IMAGING - POTENTIAL HAZARD TO PATIENT WITH EYE PROSTHESIS

被引:14
作者
YUH, WTC
HANIGAN, MT
NERAD, JA
EHRHARDT, JC
CARTER, KD
KARDON, RH
SHELLOCK, FG
机构
[1] Department of Radiology, University of Iowa College of Medicine, Iowa City, Iowa, 52242
[2] Department of Ophthalmology, University of Iowa College of Medicine, Iowa City, Iowa, 52242
[3] Department of Radiology, Cedars-Sinai Medical Center, Los Angeles
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1991年 / 1卷 / 06期
关键词
EYE; MAGNETIC RESONANCE (MR); SAFETY; PROSTHESES;
D O I
10.1002/jmri.1880010617
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Metallic implants or prostheses can be potentially hazardous during magnetic resonance (MR) imaging because of movement or dislodgment of the foreign object. Magnetic eye implants have been reported to exhibit strong movement when placed in water and exposed to a field of 1.5 T. The authors report a case of orbital implant extrusion possibly caused by the movement of a magnetic orbital implant during MR imaging at 0.5 T.
引用
收藏
页码:711 / 713
页数:3
相关论文
共 9 条
[1]  
Shellock FG, Kanal E., SMRI Safety Committee. Policies, guidelines, and recommendations for MR imaging safety and patient management, JMRI, 1, pp. 97-101, (1991)
[2]  
Kanal E, Shellock FG, Talagala L, Safety considerations in MR imaging, Radiology, 176, pp. 593-606, (1990)
[3]  
Shellock FG, Biological effects and safety aspects of magnetic resonance imaging, Magn Reson Q, 5, pp. 243-261, (1989)
[4]  
Kelly WM, Pagle PG, Pearson A, San Diego AG, Solomon MA, Ferromagnetism of intraocular foreign body causes unilateral blindness after MR study, AJNR, 7, pp. 243-245, (1986)
[5]  
Teitelbaum GP, Yee CA, Van Horn DD, Kim HS, Colletti PM, Metallic ballistic fragments: MR imaging safety and artifacts, Radiology, 175, pp. 855-859, (1990)
[6]  
Pusey E, Lufkin RB, Brown RKJ, Et al., Magnetic resonance imaging artifacts: mechanisms and clinical significance, Radio Graphics, 6, pp. 891-911, (1986)
[7]  
Shellock FG, MR imaging of metallic implants and materials: a compilation of the literature, AJR, 151, pp. 811-814, (1988)
[8]  
Power W, Collum LMT, Magnetic resonance imaging and magnetic eye implants (letter), Lancet, 2, (1988)
[9]  
Troutman RC, Five‐year survey on use of a magnetic implant for improving cosmetic result of enucleation, Arch Ophthalmol, 52, pp. 58-62, (1954)