ABSORBED DOSE TO TECHNICIANS DUE TO INDUCED ACTIVITY IN LINEAR ACCELERATORS FOR RADIATION-THERAPY

被引:36
作者
ALMEN, A
AHLGREN, L
MATTSSON, S
机构
[1] Dept. of Radiat. Phys., Lund Univ., Malmo
关键词
D O I
10.1088/0031-9155/36/6/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Absorbed dose to the trunk and to the hands of technicians working with accelerators for radiotherapy have been measured with TL dosimeters for seven different accelerators. The contribution from induced activity in the accelerator and from radiation transmitted through the walls of the treatment room have been estimated separately. The total annual absorbed dose to the trunk and to the hands have been estimated to be 2 mGy, of which the induced activity contributes one-third (0.7 mGy). The exposure of the technicians was found to be dominated by radiation penetrating the walls of the treatment room. For one accelerator the absorbed dose rate in the treatment room was measured continuously between 0.5 min and 48 h after end of treatment. Immediately after irradiation with high-energy photons the radiation is dominated by Al-26 and Cu-62 (T1/2 = 2.3 and 9.7 min respectively) and later by radionuclides with longer half-lives, W-187 and Ni-57 (T1/2 = 24 and 36h respectively). Due to these radionuclides the radioactivity in the accelerator will build up and the technicians will therefore be irradiated every time they enter the treatment room and not only directly after a treatment with high-energy photons.
引用
收藏
页码:815 / 822
页数:8
相关论文
共 7 条
  • [1] INDUCED ACTIVITY IN A HIGH-ENERGY LINEAR-ACCELERATOR
    AHLGREN, L
    OLSSON, LE
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (03) : 351 - 354
  • [2] LEDERER LP, 1978, TABLE ISOTOPES, P1
  • [3] DOSE TO RADIOTHERAPY TECHNOLOGIST FROM AIR ACTIVATION
    MCGINLEY, PH
    WRIGHT, BA
    MEDING, CJ
    [J]. MEDICAL PHYSICS, 1984, 11 (06) : 855 - 858
  • [4] RADIATION PROTECTION ASPECTS OF A NEW HIGH-ENERGY LINEAR-ACCELERATOR
    OBRIEN, P
    MICHAELS, HB
    GILLIES, B
    ALDRICH, JE
    ANDREW, JW
    [J]. MEDICAL PHYSICS, 1985, 12 (01) : 101 - 107
  • [5] A RADIATION SAFETY SURVEY ON A CLINAC-20 LINEAR-ACCELERATOR
    POWELL, NL
    NEWING, A
    BULLEN, MA
    SIMS, C
    LEATON, SF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (06) : 707 - 718
  • [6] 1980, ACTA RADIOL ONCOL, V19, P58
  • [7] 1988, SOURCES EFFECTS RISK, P10