摘要
Further developments in the measurement of chemical solutions are reported. The problem of calibration of a system for measurements of the concentration of methanol in water is described. The limitations of polynomial and inverse-polynomial models used for this purpose are demonstrated. The cubic-spline interpolation is chosen and examined as a flexible means of calibration. A method for experimental validation of the results of calibration is proposed as well. © 1990 IEEE