TRIPLE-FREQUENCY METHOD FOR MEASURING BLOOD IMPEDANCE

被引:47
作者
ZHAO, TX
JACOBSON, B
RIBBE, T
机构
[1] Department of Medical Engineering, Karolinska Institute, Novum Lluddinge University Hospital
关键词
D O I
10.1088/0967-3334/14/2/006
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A method was developed for simultaneously measuring plasma resistance, cell interior resistance and cell membrane capacitance of blood. A three-element circuit model for simulating the impedance of blood was evaluated by comparing measured and calculated impedance frequency curves and was found feasible. The values of the three elements of the model could be determined by measuring the impedance amplitudes at three different frequencies. The cell membrane capacitance (C(m)) was independent of temperature in the range 25-40-degrees-C and was significantly higher for blood with ethylenediaminetetraacetic acid (EDTA) as anticoagulant than that for blood with sodium citrate and acid citrate dextrose (ACD). The C(m) value of blood from patients with high erythrocyte sedimentation rate (ESR) was found to be significantly higher than that of blood from patients with low ESR. This suggests that the membrane capacitance might be a potential parameter for clinical applications.
引用
收藏
页码:145 / 156
页数:12
相关论文
共 25 条
[1]  
Cole K. S., 1968, MEMBRANES IONS IMPUL
[2]  
CRILE GW, 1925, J CANCER RES, V9, P3898
[3]   EFFECT OF FREQUENCY CHANGES ON ELECTRICAL CONDUCTANCE OF MOVING AND STATIONARY BLOOD [J].
FREWER, RA .
MEDICAL & BIOLOGICAL ENGINEERING, 1972, 10 (06) :734-741
[4]   The electric resistance and capacity of blood for frequencies between 800 and 4 1/2 million cycles. [J].
Fricke, H ;
Morse, S .
JOURNAL OF GENERAL PHYSIOLOGY, 1925, 9 (02) :153-167
[5]   The electric capacity of suspensions with special reference to blood. [J].
Fricke, H .
JOURNAL OF GENERAL PHYSIOLOGY, 1925, 9 (02) :137-152
[6]   A mathematical treatment of the electric conductivity and capacity of disperse systems II. The capacity of a suspension of conducting spheroids surrounded by a non-conducting membrane for a current of low frequency [J].
Fricke, H .
PHYSICAL REVIEW, 1925, 26 (05) :0678-0681
[7]   The electric impedance of hemolyzed suspensions of mammalian erythrocytes [J].
Fricke, H ;
Curtis, HJ .
JOURNAL OF GENERAL PHYSIOLOGY, 1935, 18 (06) :821-836
[8]   SPECIFIC RESISTANCE OF BLOOD AT BODY-TEMPERATURE [J].
GEDDES, LA ;
SADLER, C .
MEDICAL & BIOLOGICAL ENGINEERING, 1973, 11 (03) :336-339
[9]   SPECIFIC RESISTANCE OF BIOLOGICAL MATERIAL-A COMPENDUM OF DATA FOR BIOMEDICAL ENGINEER AND PHYSIOLOGIST [J].
GEDDES, LA ;
BAKER, LE .
MEDICAL & BIOLOGICAL ENGINEERING, 1967, 5 (03) :271-&
[10]  
GEDDES LA, 1972, CHEM INSTRUM, V4, P157