DOSE REDUCTION BY THE USE OF ERBIUM FILTRATION IN A GENERAL RADIOGRAPHIC ROOM

被引:12
作者
CRANAGE, RW
HOWARD, CJ
WELSH, AD
机构
[1] CHELTENHAM GEN HOSP,DEPT RADIOL,CHELTENHAM GL53 7AN,ENGLAND
[2] MUSGROVE PK HOSP,DEPT MED PHYS,TAUNTON TA1 5DA,SOMERSET,ENGLAND
关键词
RADIATION DOSE; FILTRATION; DOSE REDUCTION; DIAGNOSTIC RADIOLOGY;
D O I
10.1259/0007-1285-65-771-232
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The conventional aluminium filtration of a general purpose X-ray tube in a district general hospital was replaced by an erbium filter for a 2-month period. The resulting reduction in skin dose was measured for a number of examinations using thermoluminescent dosimetry. The change in effective dose equivalent was calculated from these measurements and compared with previous theoretical estimates. The use of the erbium filter had no effect on radiograph quality. It was concluded that the use of an erbium filter resulted in a worthwhile reduction in radiation dose, and was a cost-effective method of dose reduction.
引用
收藏
页码:232 / 237
页数:6
相关论文
共 16 条
  • [1] EFFECT OF ABSORPTION-EDGE FILTERS ON DIAGNOSTIC X-RAY-SPECTRA
    ATKINS, HL
    FAIRCHILD, RG
    ROBERTSON, JS
    GREENBERG, D
    [J]. RADIOLOGY, 1975, 115 (02) : 431 - 437
  • [2] PHYSICAL MEASUREMENTS OF HEAVY-METAL FILTER PERFORMANCE
    BURGESS, AE
    [J]. MEDICAL PHYSICS, 1985, 12 (02) : 225 - 228
  • [3] DOSE REDUCTION IN RADIOLOGY USING HEAVY-METAL FOILS
    CHAKERA, TMH
    FLEAY, RF
    HENSON, PW
    COLE, SM
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1982, 55 (659) : 853 - 858
  • [4] DOSE REDUCTION IN PEDIATRIC RADIOLOGY USING RARE-EARTH FILTRATION
    FLEAY, RF
    FOX, RA
    SPRAGUE, PL
    ADAMS, JP
    [J]. PEDIATRIC RADIOLOGY, 1984, 14 (05) : 332 - 334
  • [5] ENERGY-BAND SELECTION IN RADIOLOGY BY K ABSORPTION-EDGE FILTRATION
    FLEAY, RF
    [J]. AUSTRALASIAN RADIOLOGY, 1980, 24 (03): : 323 - 327
  • [6] ERBIUM FILTRATION FOR DOSE REDUCTION IN DENTAL RADIOGRAPHY
    HORNER, K
    LAWINSKI, CP
    SMITH, NJD
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1988, 61 (727) : 609 - 612
  • [7] FILTERS FOR RADIATION REDUCTION - A COMPARISON
    KOHN, ML
    GOOCH, AW
    KELLER, WS
    [J]. RADIOLOGY, 1988, 167 (01) : 255 - 257
  • [8] VALUING THE MAN-SIEVERT IN X-RAY-DIAGNOSIS
    RUSSELL, JGB
    WEBB, GAM
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1987, 60 (715) : 681 - 684
  • [9] THE INFLUENCE OF TUBE FILTRATION AND POTENTIAL ON PATIENT DOSE DURING X-RAY EXAMINATIONS
    SHRIMPTON, PC
    JONES, DG
    WALL, BF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (10) : 1205 - 1212
  • [10] EVALUATION OF THE USE OF A NIOBIUM FILTER FOR PATIENT DOSE REDUCTION IN CHEST RADIOGRAPHY
    THIERENS, H
    KUNNEN, M
    VANDERPLAETSEN, A
    SEGAERT, O
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (760) : 334 - 340