EXAMINATION OF SEPARATION EFFICIENCIES OF MERCURY-VAPOR FOR DIFFERENT GAS-LIQUID SEPARATORS IN FLOW-INJECTION COLD VAPOR ATOMIC-ABSORPTION SPECTROMETRY WITH AMALGAM PRECONCENTRATION

被引:14
作者
HANNA, CP
HAIGH, PE
TYSON, JF
MCINTOSH, S
机构
[1] UNIV MASSACHUSETTS,DEPT CHEM,AMHERST,MA 01003
[2] PERKIN ELMER CORP,DIV INORGAN ANAL,NORWALK,CT 06859
关键词
FLOW INJECTION COLD VAPOR ATOMIC ABSORPTION SPECTROMETRY; GAS LIQUID SEPARATOR; EFFICIENCY STUDY;
D O I
10.1039/ja9930800585
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A comparison has been made of the separation efficiency of three designs of gas-liquid separator when used in a flow injection (FI) manifold for the determination of Hg by cold vapour atomic absorption spectrometry. The manifold used with each device was separately optimized for maximum sensitivity. This involved studies of the effects of reagent flow rates, argon purge gas flow rate, injection time and post-injection purge time. A significant difference, with respect to both peak height and integrated signal sensitivity (by a factor of approximately 3) between the performance of a miniature design and that of two larger volume designs was obtained. No significant differences in precisions were observed. For the miniature design, the use of either tetrahydroborate or tin(II) reductant was investigated. No difference in peak height sensitivity was found, but the integrated signal sensitivity for the tetrahydroborate was 36% lower. The efficiency of separation was measured by comparison of the signal obtained from a known mass of Hg vapour, introduced via an amalgam preconcentration unit, and the signal obtained from a known mass of Hg in solution, introduced via the FI manifold and amalgam preconcentration unit. The efficiencies were found to be 101 +/- 4% and 103 +/- 6% for peak height and integrated signal, respectively.
引用
收藏
页码:585 / 588
页数:4
相关论文
共 29 条