COMPARISON OF THE LOW-CONTRAST DETECTABILITY OF A SCREEN-FILM SYSTEM AND 3RD-GENERATION COMPUTED RADIOGRAPHY

被引:24
作者
COOK, LT
INSANA, MF
MCFADDEN, MA
HALL, TJ
COX, GG
机构
[1] University of Kansas Medical Center, Department of Diagnostic Radiology, Kansas City, Kansas
关键词
CONTRAST DETAIL; COMPUTED RADIOGRAPHY; LOW-CONTRAST DETECTABILITY; OBSERVER PERFORMANCE; SCREEN-FILM;
D O I
10.1118/1.597172
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Contrast-detail (CD) analysis was used to compare the low-contrast detectability of computed radiography (CR) and screen-film (SF) as applied to the task of adult chest radiography. A phantom was constructed and imaged using the same exposure factors throughout all experiments. Within-observer variance, between-observer variance, and image sample variance were calculated and used to estimate the standard error for each experiment. The results of these CD experiments agreed with the predictions of the Rose model. Observers performed equally well for low-contrast target detection using CR and SF.
引用
收藏
页码:691 / 695
页数:5
相关论文
共 15 条
[1]   EFFICIENCY OF HUMAN VISUAL SIGNAL DISCRIMINATION [J].
BURGESS, AE ;
WAGNER, RF ;
JENNINGS, RJ ;
BARLOW, HB .
SCIENCE, 1981, 214 (4516) :93-94
[2]  
COHEN G, 1985, RECENT DEV DIGITAL I, P141
[3]  
Dainty J. C., 1974, IMAGE SCI
[4]   THRESHOLD PERCEPTION PERFORMANCE WITH COMPUTED AND SCREEN-FILM RADIOGRAPHY - IMPLICATIONS FOR CHEST RADIOGRAPHY [J].
DOBBINS, JT ;
RICE, JJ ;
BEAM, CA ;
RAVIN, CE .
RADIOLOGY, 1992, 183 (01) :179-187
[5]   HOTELLING TRACE CRITERION AND ITS CORRELATION WITH HUMAN-OBSERVER PERFORMANCE [J].
FIETE, RD ;
BARRETT, HH ;
SMITH, WE ;
MYERS, KJ .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05) :945-953
[6]   BASIC IMAGING PROPERTIES OF A COMPUTED RADIOGRAPHIC SYSTEM WITH PHOTOSTIMULABLE PHOSPHORS [J].
FUJITA, H ;
UEDA, K ;
MORISHITA, J ;
FUJIKAWA, T ;
OHTSUKA, A ;
SAI, T .
MEDICAL PHYSICS, 1989, 16 (01) :52-59
[7]  
GREEN DM, 1974, SIGNAL DETECTION THE
[8]   ULTRASOUND CONTRAST-DETAIL ANALYSIS - A PRELIMINARY-STUDY IN HUMAN OBSERVER PERFORMANCE [J].
HALL, TJ ;
INSANA, MF ;
SOLLER, NM ;
HARRISON, LA .
MEDICAL PHYSICS, 1993, 20 (01) :117-127
[9]   IMAGING PERFORMANCE OF A DIGITAL STORAGE PHOSPHOR SYSTEM [J].
HILLEN, W ;
SCHIEBEL, U ;
ZAENGEL, T .
MEDICAL PHYSICS, 1987, 14 (05) :744-751
[10]   CONTRAST-DETAIL CURVES FOR LIVER CT [J].
JUDY, PF ;
SWENSSON, RG ;
NAWFEL, RD ;
CHAN, KH ;
SELTZER, SE .
MEDICAL PHYSICS, 1992, 19 (05) :1167-1174