New Implants for the Treatment of Intertrochanteric Femur Fractures

被引:3
作者
Ricci, William M. [1 ]
机构
[1] Washington Univ, Barnes Jewish Hosp, Sch Med, Dept Orthopaed Surg, 1 Barnes Hosp Plaza,Suite 11300, St Louis, MO 63110 USA
关键词
Intertrochanteric; Fracture; Femur; New implants;
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Surgical fixation of intertrochanteric femur fractures remains the standard of care. However, the best method of surgical fixation is debatable. Fracture fixation with a compression hip screw and side plate device has long been the gold standard, but new implant designs for intramedullary nails and even for external fixation have recently challenged compression hip screws as the best method of treatment for intertrochanteric fractures. Problems related to peri-implant fracture about first-generation short intramedullary nails used for intertrochanteric fracture have, at the very least, been significantly improved by new designs. External fixation, as a result of new advances in design of external fixation pins that reduce pin tract complications and improve pin fixation strength, has recently been shown to be a reasonable alternative treatment option. Design of compression hip screw devices also continues to evolve. Plate constructs that provide biaxial dynamic compression, lag screws with improved purchase, devices with multiple lag screws, and those designed for percutaneous insertion are among the many recent innovations to compression hip screw design. Details and comparative results of these new implants, as well as the application of injectable materials for the treatment of intertrochanteric femur fractures, are reviewed in this article.
引用
收藏
页码:222 / 231
页数:10
相关论文
共 35 条
[1]
Prospective randomized controlled trial of an intramedullary nail versus dynamic screw and plate for intertrochanteric fractures of the femur [J].
Adams, CI ;
Robinson, CM ;
Court-Brown, CM ;
McQueen, MM .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2001, 15 (06) :394-400
[2]
Ahrengart L, 2002, CLIN ORTHOP RELAT R, P209
[3]
APEL DM, 1989, CLIN ORTHOP RELAT R, P156
[4]
Implant-related complications in the treatment of unstable intertrochanteric fractures:: meta-analysis of dynamic screw-plate versus dynamic screw-intramedullary nail devices [J].
Audigé, L ;
Hanson, B ;
Swiontkowski, MF .
INTERNATIONAL ORTHOPAEDICS, 2003, 27 (04) :197-203
[5]
THE EFFECT OF ADJUNCTIVE METHYLMETHACRYLATE ON FAILURES OF FIXATION AND FUNCTION IN PATIENTS WITH INTERTROCHANTERIC FRACTURES AND OSTEOPOROSIS [J].
BARTUCCI, EJ ;
GONZALEZ, MH ;
COOPERMAN, DR ;
FREEDBERG, HI ;
BARMADA, R ;
LAROS, GS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (07) :1094-1107
[6]
Baumgaertner MR, 1998, CLIN ORTHOP RELAT R, P87
[7]
THE VALUE OF THE TIP-APEX DISTANCE IN PREDICTING FAILURE OF FIXATION OF PERITROCHANTERIC FRACTURES OF THE HIP [J].
BAUMGAERTNER, MR ;
CURTIN, SL ;
LINDSKOG, DM ;
KEGGI, JM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (07) :1058-1064
[8]
Biomechanical evaluation of a new type of hip compression screw with retractable talons [J].
Bramlet, DG ;
Wheeler, D .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2003, 17 (09) :618-624
[9]
FIXATION OF INTERTROCHANTERIC FRACTURES OF THE FEMUR - A RANDOMIZED PROSPECTIVE COMPARISON OF THE GAMMA-NAIL AND THE DYNAMIC HIP SCREW [J].
BRIDLE, SH ;
PATEL, AD ;
BIRCHER, M ;
CALVERT, PT .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (02) :330-334