CONTINUOUS INTRAOPERATIVE EXTERNAL MONITORING OF PERFUSATE LEAK USING I-131 HUMAN SERUM-ALBUMIN DURING ISOLATED PERFUSION OF THE LIVER AND LIMBS

被引:34
作者
BARKER, WC [1 ]
ANDRICH, MP [1 ]
ALEXANDER, HR [1 ]
FRAKER, DL [1 ]
机构
[1] NCI,SURG BRANCH,BETHESDA,MD 20892
来源
EUROPEAN JOURNAL OF NUCLEAR MEDICINE | 1995年 / 22卷 / 11期
关键词
PERFUSION; REGIONAL; TUMOR NECROSIS FACTOR; IODINE RADIOISOTOPES; INTRAOPERATIVE MONITORING;
D O I
10.1007/BF00801607
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Regional isolated perfusion using tumor necrosis factor (TNF) shows significant promise for treatment of cancer which is limited to limbs or organs. The high toxicity of TNF requires very sensitive real time monitoring of leakage in order to avoid serious patient complications, Human serum albumin labeled with iodine-131 is used with an externally mounted and collimated NaI(Tl) detector to track the leakage of blood from the isolated perfusion blood circuit into the general systemic vascular space. Blood activity levels measured using the monitor demonstrated a very good correlation with blood serum samples taken concurrently with external monitoring. External monitoring can reduce the risks of perfusion leakage intraoperatively with the precision necessary to safely perform isolated perfusion using TNF.
引用
收藏
页码:1242 / 1248
页数:7
相关论文
共 18 条