OPTIMIZING BLOOD-VESSEL CONTRAST IN FAST 3-DIMENSIONAL MRI

被引:74
作者
HAACKE, EM
MASARYK, TJ
WIELOPOLSKI, PA
ZYPMAN, FR
TKACH, JA
AMARTUR, S
MITCHELL, J
CLAMPITT, M
PASCHAL, C
机构
[1] CASE WESTERN RESERVE UNIV,DEPT PHYS,CLEVELAND,OH 44106
[2] SIEMENS MED SYST INC,S ISELIN,NJ 08830
[3] CASE WESTERN RESERVE UNIV,DEPT BIOMED ENGN,CLEVELAND,OH 44106
[4] CASE WESTERN RESERVE UNIV,DEPT ELECT ENGN & APPL PHYS,CLEVELAND,OH 44106
[5] GE MED SYST,MILWAUKEE,WI 53201
关键词
D O I
10.1002/mrm.1910140207
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Magnetic resonance angiography has matured to the point where clinically useful images can be acquired in half an hour or less. In this paper, the role of 3D imaging techniques is primarily considered. Specifically, the optimal imaging parameters, sequences, and reconstruction techniques are evaluated for moving spins. A variant of FISP known as ROAST with low flip angles, short repeat times, and a thick slab has been found to yield the best 3D survey scan of the cranial vessels with roughly 1 × 1 × I‐mm3 resolution in each of the processed images (slices). For the faster flowing carotids, a sagittal scout with as short a TE as possible is required to avoid spin dephasing. Localization is accomplished in both cases by acquiring thin slab 3D, thin partition, larger flip angle, longer repeat time FLASH sequences. Different choices of dephaseirephase sequences and directions are also reviewed. These choices are discussed from a practical and theoretical perspective. In particular, improvements in contrast and resolution are evaluated using half‐Fourier, 512 acquisition, small fields of view and constrained reconstruction for both rephased gradient echo sequences and dephased thin slice long TR spin‐echo sequences. A resolution of 0.5‐0.75 mm is recommended to obtain sufficient image quality for consistent clinical interpretation of stenoses and vessel abnormalities. © 1990 Academic Press, Inc. Copyright © 1990 Wiley‐Liss, Inc., A Wiley Company
引用
收藏
页码:202 / 221
页数:20
相关论文
共 19 条
[1]   MR ANGIOGRAPHY OF PERIPHERAL, CAROTID, AND CORONARY-ARTERIES [J].
ALFIDI, RJ ;
MASARYK, TJ ;
HAACKE, EM ;
LENZ, GW ;
ROSS, JS ;
MODIC, MT ;
NELSON, AD ;
LIPUMA, JP ;
COHEN, AM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1987, 149 (06) :1097-1109
[2]   3-DIMENSIONAL PHASE-CONTRAST ANGIOGRAPHY [J].
DUMOULIN, CL ;
SOUZA, SP ;
WALKER, MF ;
WAGLE, W .
MAGNETIC RESONANCE IN MEDICINE, 1989, 9 (01) :139-149
[3]   INTRACEREBRAL ARTERIOVENOUS-MALFORMATIONS - EVALUATION WITH SELECTIVE MR ANGIOGRAPHY AND VENOGRAPHY [J].
EDELMAN, RR ;
WENTZ, KU ;
MATTLE, HP ;
OREILLY, GV ;
CANDIA, G ;
CHENG, L ;
BIN, Z ;
KJELLBERG, RN ;
DAVIS, KR .
RADIOLOGY, 1989, 173 (03) :831-837
[4]   NUCLEAR MAGNETIC-RESONANCE SIGNAL FROM FLOWING NUCLEI IN RAPID IMAGING USING GRADIENT ECHOES [J].
GAO, JH ;
HOLLAND, SK ;
GORE, JC .
MEDICAL PHYSICS, 1988, 15 (06) :809-814
[5]   REDUCTION OF T2-STAR DEPHASING IN GRADIENT FIELD ECHO IMAGING [J].
HAACKE, EM ;
TKACH, JA ;
PARRISH, TB .
RADIOLOGY, 1989, 170 (02) :457-462
[6]   IMPROVING MR IMAGE QUALITY IN THE PRESENCE OF MOTION BY USING REPHASING GRADIENTS [J].
HAACKE, EM ;
LENZ, GW .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1987, 148 (06) :1251-1258
[7]   CONSTRAINED RECONSTRUCTION - A SUPERRESOLUTION, OPTIMAL SIGNAL-TO-NOISE ALTERNATIVE TO THE FOURIER-TRANSFORM IN MAGNETIC-RESONANCE IMAGING [J].
HAACKE, EM ;
LIANG, ZP ;
IZEN, SH .
MEDICAL PHYSICS, 1989, 16 (03) :388-397
[8]  
HAACKE EM, IN PRESS MAGNETIC RE
[9]  
HAACKE EM, 1989, 8TH ANN SMRM M, P363
[10]  
HAACKE EM, IN PRESS RADIOLOGY