DETERMINATION OF CARBONYL-COMPOUNDS IN AIR BY HPLC USING ONLINE ANALYZED MICROCARTRIDGES, FLUORESCENCE AND CHEMILUMINESCENCE DETECTION

被引:40
作者
NONDEK, L
MILOFSKY, RE
BIRKS, JW
机构
[1] UNIV COLORADO,DEPT CHEM & BIOCHEM,CAMPUS BOX 216,BOULDER,CO 80309
[2] UNIV COLORADO,COOPERAT INST RES ENVIRONM SCI,BOULDER,CO 80309
关键词
COLUMN LIQUID CHROMATOGRAPHY; AIR ANALYSIS; DANSYLHYDRAZONES OF CARBONYL COMPOUNDS; CHEMILUMINESCENCE; FLUORESCENCE;
D O I
10.1007/BF02262463
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Sensitive and selective detection of dansylhydrazones of atmospheric carbonyl compounds (aldehydes and ketones) can be achieved using high performance liquid chromatography (HPLC) with fluorescence or chemiluminescence detection. The carbonyl compounds are derivatized by drawing air through small glass cartridges packed with porous glass particles impregnated with dansylhydrazine. After sampling, the contents of the cartridges are analyzed on-line by using a small plug of water (200-mu-L) to transfer and focus the hydrazone derivatives at the head of a HPLC column. Greatly increased sensitivity over traditional methods derives from 1) analysis of the entire contents of the sampling cartridge, and 2) detection by fluorescence or peroxyoxalate chemilum-inescence. Results are compared for photo-initiated and H2O2-initiated peroxyoxalate chemiluminescence. This novel and practical system enables the detection of sub-ppbv concentrations of formaldehyde, acetaldehyde, acetone and higher carbonyls in air using relatively short sampling times.
引用
收藏
页码:33 / 39
页数:7
相关论文
共 36 条
[1]   2,4-DINITROPHENYLHYDRAZINE-COATED SILICA-GEL CARTRIDGE METHOD FOR DETERMINATION OF FORMALDEHYDE IN AIR - IDENTIFICATION OF AN OZONE INTERFERENCE [J].
ARNTS, RR ;
TEJADA, SB .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1989, 23 (11) :1428-1430
[2]   CHEMI-LUMINESCENCE FROM REACTIONS OF BIS(PENTACHLOROPHENYL)OXALATE, HYDROGEN-PEROXIDE AND FLUORESCENT COMPOUNDS - KINETICS AND MECHANISM [J].
CATHERALL, CLR ;
PALMER, TF ;
CUNDALL, RB .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS II, 1984, 80 :823-836
[3]  
CHANDROSS EA, 1963, TETRAHEDRON LETT, P761
[4]   SAMPLING OF ATMOSPHERIC CARBONYLS WITH SMALL DNPH-COATED C18 CARTRIDGES AND LIQUID-CHROMATOGRAPHY ANALYSIS WITH DIODE-ARRAY DETECTION [J].
DRUZIK, CM ;
GROSJEAN, D ;
VANNESTE, A ;
PARMAR, SS .
INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 1990, 38 (04) :495-512
[5]   AN IMPROVED HPLC METHOD FOR CARBONYL COMPOUND SPECIATION IN THE ATMOSPHERIC LIQUID-PHASE [J].
FACCHINI, MC ;
CHIAVARI, G ;
FUZZI, S .
CHEMOSPHERE, 1986, 15 (06) :667-674
[6]  
Givens R.S., 1989, CHEMILUMINESCENCE PH, P125
[7]   DETERMINATION OF ALDEHYDES AND KETONES IN NATURAL-GAS COMBUSTION IN THE PPB RANGE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
GOTZE, HJ ;
HARKE, S .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1989, 335 (03) :286-288
[8]   SOME NEW ASPECTS OF A HPLC-METHOD FOR THE DETERMINATION OF TRACES OF FORMALDEHYDE IN AIR [J].
GROMPING, A ;
CAMMANN, K .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1989, 335 (07) :796-801
[9]   COLLECTION EFFICIENCIES OF CARTRIDGES AND MICRO-IMPINGERS FOR SAMPLING OF ALDEHYDES IN AIR AS 2,4-DINITROPHENYLHYDRAZONES [J].
GROSJEAN, D ;
FUNG, K .
ANALYTICAL CHEMISTRY, 1982, 54 (07) :1221-1224
[10]   FORMALDEHYDE AND OTHER CARBONYLS IN LOS-ANGELES AMBIENT AIR [J].
GROSJEAN, D .