PRELIMINARY EVALUATION OF A DIGITAL SYSTEM FOR ROTATIONAL PANORAMIC RADIOGRAPHY

被引:10
作者
DOVE, SB [1 ]
MCDAVID, WD [1 ]
WELANDER, U [1 ]
TRONJE, G [1 ]
机构
[1] KAROLINSKA INST,DEPT ORAL RADIOL,S-10401 STOCKHOLM 60,SWEDEN
来源
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS | 1992年 / 73卷 / 05期
关键词
D O I
10.1016/0030-4220(92)90111-3
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
A prototype system for direct digital panoramic radiography has been evaluated with respect to density, contrast, magnification, distortion, resolution, and overall image quality. Density and contrast depend on detector calibration and may be modified by the display system or by digital processing of the captured image. Variation of magnification in the horizontal and vertical dimensions gives rise to distortion phenomena that are identical to those encountered in film-based systems. Resolution in the vertical dimension is determined by the pitch of the detector elements. In the horizontal dimension, resolution is limited by the effective width of the detector elements. To evaluate the clinical acceptability of the images, radiologists and general practice residents were asked to assess the perceptibility of important radiographic landmarks in film-based and digital images of both a radiographic phantom and a patient. The digital system performed on a par with film in the representation of normal morphologic structures of the clinical human subject whereas more differences were apparent in the phantom images. The general practice residents consistently rated the digital images higher than their radiologist counterparts did. No consistent trends were found to indicate any inherent deficiencies of the digital system in the depiction of any one area. The results indicate the promise of direct digital acquisition as a method of panoramic imaging.
引用
收藏
页码:623 / 632
页数:10
相关论文
共 12 条
[1]  
DAHLIN H, 1978, ACTA RADIOL S, P356
[2]   FASTER SCREEN-FILM COMBINATIONS FOR CEPHALOMETRIC RADIOGRAPHY [J].
HURLBURT, CE .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1981, 52 (06) :661-665
[3]  
LEITE LP, 1990, J DENT RES, V69, P344
[4]   ELECTRONIC SYSTEM FOR DIGITAL ACQUISITION OF ROTATIONAL PANORAMIC RADIOGRAPHS [J].
MCDAVID, WD ;
DOVE, SB ;
WELANDER, U ;
TRONJE, G .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1991, 71 (04) :499-502
[5]  
MCDAVID WD, 1985, DENTOMAXILLOFAC RA S, V8, P13
[6]  
MUNIER B, 1987, SOLID STATE 1024 PIX, P2
[7]   350 KVP CHEST RADIOGRAPHY - REVIEW AND COMPARISON WITH 120 KVP [J].
PROTO, AV ;
LANE, EJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1978, 130 (05) :859-866
[8]  
SICKLES EA, 1977, APPLICATION OPTICAL, V6, P30
[9]   DIGITAL RADIOGRAPHY OF THE CHEST - DESIGN-FEATURES AND CONSIDERATIONS FOR A PROTOTYPE UNIT [J].
TESIC, MM ;
MATTSON, RA ;
BARNES, GT ;
SONES, RA ;
STICKNEY, JB .
RADIOLOGY, 1983, 148 (01) :259-264
[10]  
Tronje G., 1985, DENTOMAXILLOFAC RA S, V8, P29