CALIBRATION OF A TSI-MODEL-3025 ULTRAFINE CONDENSATION PARTICLE COUNTER

被引:77
作者
KESTEN, J
REINEKING, A
PORSTENDORFER, J
机构
[1] Isotope Laboratory for Biological and Medical Research, University of Göttingen, D-3400, Göttingen
关键词
D O I
10.1080/02786829108959517
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The registration efficiency of the TSI model 3025 ultrafine condensation particle counter for Ag and NaCl particles of between 2 and 20 nm in diameter was determined. Taking into account the different shapes of the input aerosol size distributions entering the differential mobility analyzer (DMA) and the transfer function of the DMA, the counting efficiencies of condensation nucleus counters (CNC) for monodisperse Ag and NaCl particles were estimated. In addition, the dependence of the CNC registration efficiency on the particle concentration was investigated.