TOLERANCES IN SETUP AND DOSIMETRIC ERRORS IN THE RADIATION TREATMENT OF BREAST-CANCER

被引:34
作者
DAS, IJ [1 ]
CHENG, CW [1 ]
FOSMIRE, H [1 ]
KASE, KR [1 ]
FITZGERALD, TJ [1 ]
机构
[1] UNIV ARIZONA,ARIZONA HLTH SCI CTR,TUCSON,AZ 85724
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1993年 / 26卷 / 05期
关键词
BREAST CANCER; RADIATION TREATMENT; SETUP ERROR; DOSE VOLUME HISTOGRAM; SIMULATION;
D O I
10.1016/0360-3016(93)90505-P
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Treatment failure in radiation therapy, as well as unexpected complications, can be associated with set up changes or variations that can cause deviations from the prescribed radiation dose distribution both inside and outside the target volume. The effect of various deviations from the planned setup on the delivery of the prescribed radiation dose to the desired treatment volume was studied. Methods and Materials: Adding a second simulation was investigated as means of minimizing setup changes on treatment. The first simulation was used for planning the treatment and the second simulation was essentially a mock treatment. Dosimetric evaluations based on dose volume histograms were analyzed for each deviation in the setup. Results: In 95% of the patients, the frequency of the changes in the setup parameters between the second simulation and the treatment setup were reduced significantly from the changes that occurred between the first simulation and the second simulation. The changes in isocenter coordinates up to +/-1.0 cm have minimal effects (+/-2%) on the dose distributions. Gantry angle variations up to +/-4-degrees produce a change of less than +/-5% in the dose distribution within the target volume. However, this angular variation resulted in additional tissue irradiation outside of the desired treatment field (about 10 cm3 for a large patient). A gantry angle variation of +/-6-degrees can change the volume of tissue that receives the prescribed dose by at least +/-10%. In addition, such a change can increase the volume of tissue outside the desired treatment field that is irradiated. Conclusion: It is concluded that individually, deviations in one of the parameters from the planned setup of +/-1.0 cm in isocenter position and +/-4-degrees in gantry angle do not produce significant deviations from the planned dose distribution. However, a significant change in dose distribution is observed if the setup parameters are concurrently changed. A second simulation may minimize the deviations of the treatment setup from the planned setup and maximize the precision in dose delivery to the target volume.
引用
收藏
页码:883 / 890
页数:8
相关论文
共 20 条
[1]   WEEKLY LOCALIZATION FILMS AND DETECTION OF FIELD PLACEMENT ERRORS [J].
BYHARDT, RW ;
COX, JD ;
HORNBURG, A ;
LIERMANN, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1978, 4 (9-10) :881-887
[2]   3-DIMENSIONAL PHOTON DOSE DISTRIBUTIONS WITH AND WITHOUT LUNG CORRECTIONS FOR TANGENTIAL BREAST INTACT TREATMENTS [J].
CHIN, LM ;
CHENG, CW ;
SIDDON, RL ;
RICE, RK ;
MIJNHEER, BJ ;
HARRIS, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (06) :1327-1335
[3]   5-YEAR RESULTS OF A RANDOMIZED CLINICAL-TRIAL COMPARING TOTAL MASTECTOMY AND SEGMENTAL MASTECTOMY WITH OR WITHOUT RADIATION IN THE TREATMENT OF BREAST-CANCER [J].
FISHER, B ;
BAUER, M ;
MARGOLESE, R ;
POISSON, R ;
PILCH, Y ;
REDMOND, C ;
FISHER, E ;
WOLMARK, N ;
DEUTSCH, M ;
MONTAGUE, E ;
SAFFER, E ;
WICKERHAM, L ;
LERNER, H ;
GLASS, A ;
SHIBATA, H ;
DECKERS, P ;
KETCHAM, A ;
OISHI, R ;
RUSSELL, I .
NEW ENGLAND JOURNAL OF MEDICINE, 1985, 312 (11) :665-673
[4]   PRIMARY BREAST IRRADIATION IN LARGE-BREASTED OR HEAVY WOMEN - ANALYSIS OF COSMETIC OUTCOME [J].
GRAY, JR ;
MCCORMICK, B ;
COX, L ;
YAHALOM, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (02) :347-354
[5]  
KHAN FM, 1973, RADIOLOGY, V149, P318
[6]   RADIATION PNEUMONITIS IN BREAST-CANCER PATIENTS TREATED WITH CONSERVATIVE SURGERY AND RADIATION-THERAPY [J].
LINGOS, TI ;
RECHT, A ;
VICINI, F ;
ABNER, A ;
SILVER, B ;
HARRIS, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (02) :355-360
[7]   SCATTER DOSE FROM TANGENTIAL BREAST IRRADIATION TO THE UNINVOLVED BREAST [J].
MULLERRUNKEL, R ;
KALOKHE, UP .
RADIOLOGY, 1990, 175 (03) :873-876
[8]   ACCURACY OF RADIATION-FIELD ALIGNMENT IN CLINICAL-PRACTICE [J].
RABINOWITZ, I ;
BROOMBERG, J ;
GOITEIN, M ;
MCCARTHY, K ;
LEONG, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (10) :1857-1867
[9]   IMPROVED METHODS FOR DETERMINATION OF VARIABILITY IN PATIENT POSITIONING FOR RADIATION-THERAPY USING SIMULATION AND SERIAL PORTAL FILM MEASUREMENTS [J].
ROSENTHAL, SA ;
GALVIN, JM ;
GOLDWEIN, JW ;
SMITH, AR ;
BLITZER, PH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 23 (03) :621-625
[10]   RADIATION PNEUMONITIS IN PATIENTS TREATED FOR BREAST-CANCER [J].
ROTHWELL, RI ;
KELLY, SA ;
JOSLIN, CAF .
RADIOTHERAPY AND ONCOLOGY, 1985, 4 (01) :9-14