USE OF A PACKED PROGRAMMED-TEMPERATURE VAPORIZER INJECTOR IN THE SOLVENT ELIMINATION MODE FOR THE DETERMINATION OF FATTY-ACID METHYL-ESTERS BY GAS-CHROMATOGRAPHY

被引:17
作者
MEDINA, I [1 ]
LINARES, F [1 ]
GARRIDO, JL [1 ]
机构
[1] CTR INVEST MARINAS,VILAGARCIA AROUSA,SPAIN
关键词
D O I
10.1016/0021-9673(94)85089-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
It is shown that when using a programmed-temperature vaporizer as the injection device for the gas chromatographic determination of fatty acid methyl esters in the solvent elimination mode, losses of medium-chain solutes during the solvent elimination step can be avoided by packing the glass liner of the injector with materials providing adequate retention characteristics. Silanized glass-wool proved to be a good material for this purpose. Other parameters such as solvent elimination temperature, nature of the solvent and solute vaporization temperature were also studied in order to achieve optimum injection conditions.
引用
收藏
页码:472 / 476
页数:5
相关论文
共 12 条
  • [1] ANALYSIS OF FATTY-ACID METHYL-ESTERS WITH HIGH-ACCURACY AND RELIABILITY .6. RAPID ANALYSIS BY SPLIT INJECTION CAPILLARY GAS-LIQUID-CHROMATOGRAPHY
    BANNON, CD
    CRASKE, JD
    FELDER, DL
    GARLAND, IJ
    NORMAN, LM
    [J]. JOURNAL OF CHROMATOGRAPHY, 1987, 407 : 231 - 241
  • [2] Christie W.W., 1989, GAS CHROMATOGRAPHY L
  • [3] GAS-CHROMATOGRAPHIC ANALYSIS OF FATTY-ACID METHYL-ESTERS - AVOIDING DISCRIMINATION BY PROGRAMMED TEMPERATURE VAPORIZING INJECTION
    EDER, K
    REICHLMAYRLAIS, AM
    KIRCHGESSNER, M
    [J]. JOURNAL OF CHROMATOGRAPHY, 1991, 588 (1-2): : 265 - 272
  • [4] SAMPLING OF VOLATILE COMPONENTS USING A PTV IN THE SOLVENT SPLIT MODE
    HERRAIZ, M
    REGLERO, G
    LOYOLA, E
    HERRAIZ, T
    [J]. JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1987, 10 (11): : 598 - 602
  • [5] JOSEPH JD, 1992, J AOAC INT, V75, P488
  • [6] LEPAGE G, 1986, J LIPID RES, V27, P114
  • [7] EVALUATION OF A PROGRAMMED TEMPERATURE VAPORIZER FOR QUANTITATIVE-ANALYSIS BY SPLIT INJECTION
    LOYOLA, E
    HERRAIZ, M
    REGLERO, G
    MARTINALVAREZ, P
    [J]. JOURNAL OF CHROMATOGRAPHY, 1987, 398 : 53 - 61
  • [8] CAPILLARY GAS-CHROMATOGRAPHIC PROFILING OF TOTAL LONG-CHAIN FATTY-ACIDS AND CHOLESTEROL IN BIOLOGICAL-MATERIALS
    MUSKIET, FAJ
    VANDOORMAAL, JJ
    MARTINI, IA
    WOLTHERS, BG
    VANDERSLIK, W
    [J]. JOURNAL OF CHROMATOGRAPHY, 1983, 278 (02): : 231 - 244
  • [9] A UNIVERSAL SAMPLE INJECTION SYSTEM FOR CAPILLARY COLUMN GC USING A PROGRAMMED TEMPERATURE VAPORIZER (PTV)
    POY, F
    VISANI, S
    TERROSI, F
    [J]. JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1982, 5 (07): : 355 - 359
  • [10] SAMPLING IN CAPILLARY GAS-CHROMATOGRAPHY - A COMPARISON BETWEEN THE SPLIT AND THE PTV-SPLIT PROCEDURES
    REGLERO, G
    HERRAIZ, M
    CABEZUDO, MD
    [J]. CHROMATOGRAPHIA, 1986, 22 (7-12) : 333 - 336