EFFECT OF SEPARATION CONDITIONS ON AUTOMATED ISOELECTRIC-FOCUSING OF CARBOHYDRATE-DEFICIENT TRANSFERRIN AND OTHER HUMAN ISOTRANSFERRINS USING THE PHASTSYSTEM

被引:33
作者
HACKLER, R
ARNDT, T
KLEINE, TO
GRESSNER, AM
机构
[1] UNIV MARBURG,DEPT CLIN CHEM,D-35033 MARBURG,GERMANY
[2] UNIV MARBURG,CENT LAB,D-35033 MARBURG,GERMANY
[3] UNIV MARBURG,DEPT NEUROCHEM,CTR NERVOUS DIS,D-35033 MARBURG,GERMANY
关键词
D O I
10.1006/abio.1995.1475
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
To investigate the effect of automated isoelectric focusing conditions in the PhastSystem, e.g., the point of sample application, prerun and separation times, and minimized gels on isotransferrin band pattern, human sera were analyzed with native transferrin iron load, after iron saturation or iron depletion in vitro. Varying the focusing conditions we found (i) Point of sample application (anode, middle of the gel, cathode) strongly affected transferrin iron loss. It was greatest at the anode and least at the cathode. (ii) Without pre run, distinct transferrin iron loss also occurred, A short prerun time prevented iron loss, but increasing it did not improve transferrin iron load stability as stated by others. (iii) An inappropriately long separation time inevitably yielded iron loss. In conclusion, inappropriate isoelectric focusing conditions strongly affect iron load stability of isotransferrins (obviously via low pH within the gel), resulting in transferrin iron release and cofocusing of isotransferrins with different sialic acid or iron contents. For determination of carbohydrate-deficient transferrin, such conditions resulted in overestimation of the marker of chronic alcohol abuse. Our findings may be of guiding importance for isoelectric focusing of protein-ligand complexes. We recommend the procedure described for development of isoelectric focusing of protein-ligand complexes. (C) 1995 Academic Press, Inc.
引用
收藏
页码:281 / 289
页数:9
相关论文
共 29 条
[1]  
ALLEN RC, 1977, ELECTROFOCUSING ISOT, P255
[2]   DEMONSTRATION OF HUMAN PRE-ALBUMIN BY DOUBLE ONE-DIMENSIONAL SLAB GEL-ELECTROPHORESIS [J].
ALTLAND, K ;
RAUH, S ;
HACKLER, R .
ELECTROPHORESIS, 1981, 2 (03) :148-155
[3]   DOUBLE ONE-DIMENSIONAL ELECTROPHORESIS OF HUMAN-SERUM TRANSFERRIN - A NEW HIGH-RESOLUTION SCREENING METHOD FOR GENETICALLY-DETERMINED VARIATION [J].
ALTLAND, K ;
HACKLER, R ;
KNOCHE, W .
HUMAN GENETICS, 1980, 54 (02) :221-231
[4]  
ALTLAND K, 1977, ELECTROFOCUSING ISOT, P295
[5]  
ARNDT T, 1994, MED WELT, V45, P247
[7]  
BATES GW, 1967, J BIOL CHEM, V242, P2816
[8]   A NEW APPROACH TO QUANTITATE CARBOHYDRATE-DEFICIENT TRANSFERRIN ISOFORMS IN ALCOHOL ABUSERS - PARTIAL IRON SATURATION IN ISOELECTRIC FOCUSING/IMMUNOBLOTTING AND LASER DENSITOMETRY [J].
BEAN, P ;
PETER, JB .
ALCOHOLISM-CLINICAL AND EXPERIMENTAL RESEARCH, 1993, 17 (06) :1163-1170
[9]   MICROHETEROGENEITY OF HUMAN-SERUM TRANSFERRIN - A BIOLOGICAL PHENOMENON STUDIED BY ISOELECTRIC-FOCUSING IN IMMOBILIZED PH GRADIENTS [J].
DEJONG, G ;
VANEIJK, HG .
ELECTROPHORESIS, 1988, 9 (09) :589-598
[10]   THE BIOLOGY OF TRANSFERRIN [J].
DEJONG, G ;
VANDIJK, JP ;
VANEIJK, HG .
CLINICA CHIMICA ACTA, 1990, 190 (1-2) :1-46