Successful retrieval of migrated Amplatzer septal occluder

被引:14
作者
Shirodkar, Salil [1 ]
Patil, Sushant [1 ]
Pinto, Robin [1 ]
Dalvi, Bharat [1 ]
机构
[1] Dr Balabhai Nanavati Hosp, Dept Cardiol, Mumbai, Maharashtra, India
关键词
Atrial septal defect; bioptome; device embolization; goose neck snare;
D O I
10.4103/0974-2069.64365
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Migration and embolization of the Amplatzer septal occluder (ASO) have been reported. However, there is only limited information on the methods of systematic retrieval of these devices. In this report, we describe the a case of a 4 year old girl who underwent closure of her atrial septal defect (ASD) using a 17 mm ASO. The device migrated in to the right atrium an anteroposterior plane 24 hours later with a resultant residual shunt. The device was successfully retrieved percutaneously and the ASD was closed using a 20 mm ASO.
引用
收藏
页码:83 / 86
页数:4
相关论文
共 7 条
[1]
New technique for device closure of large atrial septal defects [J].
Dalvi, BV ;
Pinto, RJ ;
Gupta, A .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 64 (01) :102-107
[2]
Use of a straight, side-hole delivery sheath for improved delivery of amplatzer ASD occluder [J].
Kutty, Shelby ;
Asnes, Jeremy D. ;
Srinath, Gowdagere ;
Preminger, Tamar J. ;
Prieto, Lourdes R. ;
Latson, Larry A. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (01) :15-20
[3]
Embolization and retrieval of the amplatzer septal occluder [J].
Levi, DS ;
Moore, JW .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 61 (04) :543-547
[4]
Masura J, 1997, CATHETER CARDIO DIAG, V42, P388, DOI 10.1002/(SICI)1097-0304(199712)42:4<388::AID-CCD7>3.0.CO
[5]
2-7
[6]
Peuster M, 2000, CATHETER CARDIO INTE, V51, P297, DOI 10.1002/1522-726X(200011)51:3<297::AID-CCD11>3.0.CO
[7]
2-O