3-DIMENSIONAL NMR MICROSCOPY - IMPROVING SNR WITH TEMPERATURE AND MICROCOILS

被引:47
作者
MCFARLAND, EW
MORTARA, A
机构
[1] Department of Nuclear Engineering, Massachusetts Institute of Technology, Cambridge, MA
[2] Department of Chemical and Nuclear Engineering, University of California, Santa Barbara, CA
关键词
NMR MICROSCOPY; SIGNAL-TO-NOISE RATIO; CRYOGENICS; PREAMPLIFIER;
D O I
10.1016/0730-725X(92)90487-K
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
It is widely held that the spatial resolution achievable by NMR microscopic imaging is limited in biological systems by diffusion to approximately 1-5-mu-m. However, these estimates were developed for specific imaging techniques and represent practical rather than fundamental limits. NMR imaging is limited by the signal-to-noise ratio (SNR). Diffusion effects on spatial resolution can be made arbitrarily small in principle by increasing the gradient strength. The exponential signal attenuation from random spin motion in a gradient, however, will reduce the signal far below the noise level when the voxel size is reduced much below 5-mu-m. Two factors can be optimized to improve the SNR: (1) the inductive linkage between microscopic samples and the detection apparatus and (2) the temperature of the rf probe. In this work, the filling factor was optimized using inductors with diameters less than 1 mm. It is furthermore shown that probe circuit cooling results in significant improvements in SNR, whereas cooling of the preamplifier is of little value when proper noise matching between the resonant circuit and preamplifier is accomplished. Using three-dimensional Fourier imaging techniques, we have obtained images of single-cell organisms with spatial resolution of approximately 6-mu-m. Practical limitations include mechanical stability of the apparatus, thermal shielding between the sample and probe, and the magnetic susceptibility of the sample.
引用
收藏
页码:279 / 288
页数:10
相关论文
共 25 条
  • [1] Mansfield, Grannel, “Diffraction” and microscopy in solids and liquids by NMR, Phys. Rev. B., 12, 9, pp. 3618-3634, (1975)
  • [2] Hinshaw, Image formation by nuclear magnetic resonance: The sensitive-point method, J. Appl. Phys., 47, 8, pp. 3709-3721, (1976)
  • [3] Hedges, Microscopic NMR imaging, Ph.D. thesis, (1984)
  • [4] McFarland, Kushmerick, Neuringer, Chemical exchange magnetic resonance, Magn. Reson. Imag., 6, pp. 507-515, (1988)
  • [5] Aguayo, Blackband, Schoeniger, Mattingly, Hinterman, Nuclear magnetic resonance of a single cell, Nature, 322, (1987)
  • [6] Meyer, Brown, Diffusion measurements by microscopic NMR imaging, J. Magn. Reson., 76, pp. 393-399, (1988)
  • [7] Zhou, Porter, Lauterbur, Voth, 3D microscopic NMR imaging with (6.37 μm)3 isotropic resolution, 8th Annual Meeting of Society of Magnetic Resonance in Medicine, (1989)
  • [8] Glover, Bowtell, Harvey, McJury, Brown, Mansfield, A high resolution NMR microscope for use at 11.7 T, 8th Annual Meeting of Society of Magnetic Resonance in Medicine, (1989)
  • [9] Johnson, Thompson, Gewalt, Hayes, Nuclear magnetic resonance imaging at microscopic resolution, J. Magn. Reson., 68, pp. 129-137, (1986)
  • [10] Eccles, Callaghan, High-resolution imaging. The NMR microscope, J. Magn. Reson., 68, pp. 393-398, (1986)