COLOR DISCRIMINATION AND ACCURACY OF BLOOD-GLUCOSE SELF-MONITORING IN TYPE-I DIABETIC-PATIENTS

被引:8
作者
SAWICKI, PT
KARSCHNY, L
STOLPE, V
WOLF, E
BERGER, M
机构
[1] UNIV DUSSELDORF,DEPT NEURO & SENSORY PHYSIOL,W-4000 DUSSELDORF 1,GERMANY
[2] UNIV DUSSELDORF,DEPT NUTR & METAB,WHO,COLLABORATING CTR DIABET,W-4000 DUSSELDORF 1,GERMANY
关键词
D O I
10.2337/diacare.14.2.135
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To determine the importance of color discrimination ability regarding accuracy in the self-monitoring of blood glucose. Research Design and Methods: Two hundred two insulin-dependent (type 1) diabetic patients performed the Fansworth-Munsell 100-hue test and a second group of 159 type 1 diabetic patients performed a new Dusseldorf 26-hue test. Error scores in both tests were compared with error scores of patients' self-monitoring of blood glucose measurements with Chemstrip Haemoglucotest 20-800 strips. Results: Color discrimination ability decreased with age, diabetes duration, and presence of retinopathy. It was independent of the degree of glycemia and accuracy of self-monitoring of blood glucose. Conclusions: Impaired color vision by itself is no reason to abandon self-monitoring of blood glucose with visually read strips.
引用
收藏
页码:135 / 137
页数:3
相关论文
共 18 条
[1]   PREDICTION OF DIABETIC-RETINOPATHY FROM CLINICAL-VARIABLES AND COLOR-VISION DATA [J].
ASPINALL, PA ;
KINNEAR, PR ;
DUNCAN, LJP ;
CLARKE, BF .
DIABETES CARE, 1983, 6 (02) :144-148
[2]   EARLY LOSS OF BLUE-SENSITIVE COLOR-VISION IN PATIENTS WITH TYPE-I DIABETES [J].
DALEY, ML ;
WATZKE, RC ;
RIDDLE, MC .
DIABETES CARE, 1987, 10 (06) :777-781
[3]  
FARNSWORTH D, 1957, FARNSWORTHMUNSELL 10
[4]   RELEVANCE OF COLOR-VISION AND DIABETIC-RETINOPATHY TO SELF-MONITORING OF BLOOD-GLUCOSE [J].
GRAHAM, K ;
KESSON, CM ;
KENNEDY, HB ;
IRELAND, JT .
BRITISH MEDICAL JOURNAL, 1980, 281 (6246) :971-973
[5]   A PRECISION INDEX FOR EVALUATION OF TECHNIQUES FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
HALLE, JP ;
LAMBERT, J ;
RENAUD, E ;
TOUPIN, A ;
LEGAULT, C ;
LALUMIERE, G .
DIABETES CARE, 1986, 9 (03) :319-320
[6]   ABNORMAL COLOR-VISION AND RELIABLE SELF-MONITORING OF BLOOD-GLUCOSE [J].
LOMBRAIL, P ;
CATHELINEAU, G ;
GERVAIS, P ;
THIBULT, N .
DIABETES CARE, 1984, 7 (04) :318-321
[7]  
MANTYJARVI M, 1989, ACTA OPHTHALMOL, V67, P113
[8]   USE AND ABUSE OF GLUCOSE REFLECTANCE METERS [J].
MOUNTIER, VM ;
SCOTT, RS ;
BEAVEN, DW .
DIABETES CARE, 1982, 5 (05) :542-544
[9]  
MUHLHAUSER I, 1983, DIABETOLOGIA, V25, P470
[10]   QUALITY-CONTROL OF SELF-MONITORING OF BLOOD-SUGAR LEVELS IN A GROUP OF UNSELECTED TYPE-I DIABETICS - IMPORTANCE OF TRAINING THE PATIENTS [J].
MUHLHAUSER, I ;
BROERMANN, C ;
BARTELS, H ;
BERG, R ;
SCHWARZ, S ;
BERGER, M .
DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1984, 109 (41) :1553-1557