PHASE-ENCODE ORDER AND ITS EFFECT ON CONTRAST AND ARTIFACT IN SINGLE-SHOT RARE SEQUENCES

被引:51
作者
MULKERN, RV
MELKI, PS
JAKAB, P
HIGUCHI, N
JOLESZ, FA
机构
[1] Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston., Massachusetts
关键词
IMAGING SEQUENCES; NMR; MR; AGENTS; MOTION;
D O I
10.1118/1.596644
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Substantial manipulation of tissue contrast can be achieved by varying the order in which phase-encode values are applied to individual echoes within a 128-echo single-shot rapid acquisition relaxation enhanced (RARE) sequence. Appropriate ordering can then permit imaging of short T2 species like muscle and white matter with single-shot RARE. For sequential phase encoding with an arbitrary initial phase-encode value, the timing of the zero phase (ZP) encoded echo is found to be analogous to the echo time (TE) of standard spin-echo sequences. This is demonstrated qualitatively with human brain images and is verified quantitatively with NiCl2 phantoms by correlating the time constant for signal decay with ZP echo time, with transverse relaxation times T2, as obtained with a 128-echo Carr-Purcell-Meiboom-Gill (CPMG) imaging sequence. Banding artifacts accompanying the discontinuous traverse through K space are experimentally demonstrated in a rectangular phantom and expressions are developed for determining the dependence of this artifact on the phase-encode gradient increments and durations, the ZP echo number, echo spacing, and T2. Simulations based on the expressions are shown to be useful for characterizing the observed "banding" artifacts perpendicular to the phase-encode direction and for predicting the extent of tissue-tissue overlap to be expected with the use of this ultrafast rf echo planar imaging method.
引用
收藏
页码:1032 / 1037
页数:6
相关论文
共 20 条
[1]   RESPIRATORY ORDERED PHASE ENCODING (ROPE) - A METHOD FOR REDUCING RESPIRATORY MOTION ARTIFACTS IN MR IMAGING [J].
BAILES, DR ;
GILDERDALE, DJ ;
BYDDER, GM ;
COLLINS, AG ;
FIRMIN, DN .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1985, 9 (04) :835-838
[2]   REAL-TIME MOVIE IMAGING FROM A SINGLE CARDIAC CYCLE BY NMR [J].
CHAPMAN, B ;
TURNER, R ;
ORDIDGE, RJ ;
DOYLE, M ;
CAWLEY, M ;
COXON, R ;
GLOVER, P ;
MANSFIELD, P .
MAGNETIC RESONANCE IN MEDICINE, 1987, 5 (03) :246-254
[3]   ANALYSIS OF T2 LIMITATIONS AND OFF-RESONANCE EFFECTS ON SPATIAL-RESOLUTION AND ARTIFACTS IN ECHO-PLANAR IMAGING [J].
FARZANEH, F ;
RIEDERER, SJ ;
PELC, NJ .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (01) :123-139
[4]   RAPID NMR IMAGING OF DYNAMIC PROCESSES USING THE FLASH TECHNIQUE [J].
FRAHM, J ;
HAASE, A ;
MATTHAEI, D .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (02) :321-327
[5]   RAPID DATA-ACQUISITION TECHNIQUE FOR NMR IMAGING BY THE PROJECTION RECONSTRUCTION METHOD [J].
HALL, LD ;
SUKUMAR, S .
JOURNAL OF MAGNETIC RESONANCE, 1984, 56 (01) :179-182
[6]   RAPID FOURIER-IMAGING USING STEADY-STATE FREE PRECESSION [J].
HAWKES, RC ;
PATZ, S .
MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (01) :9-23
[7]   RARE IMAGING - A FAST IMAGING METHOD FOR CLINICAL MR [J].
HENNIG, J ;
NAUERTH, A ;
FRIEDBURG, H .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (06) :823-833
[8]  
HENNIG J, 1986, J COMPUT ASSIST TOMO, V10, P375
[9]   MULTIECHO IMAGING SEQUENCES WITH LOW REFOCUSING FLIP ANGLES [J].
HENNIG, J .
JOURNAL OF MAGNETIC RESONANCE, 1988, 78 (03) :397-407
[10]   FAST 3-DIMENSIONAL IMAGING OF CEREBROSPINAL-FLUID [J].
HENNIG, J ;
FRIEDBURG, H ;
OTT, D .
MAGNETIC RESONANCE IN MEDICINE, 1987, 5 (04) :380-383