PERFORMANCE OF A HYDROPHOBIC HEAT AND MOISTURE EXCHANGER AT DIFFERENT AMBIENT-TEMPERATURES

被引:7
作者
CROCI, M [1 ]
ELENA, A [1 ]
SOLCA, M [1 ]
机构
[1] UNIV MILAN,IST ANESTESIOL & REANIMAZ,I-20122 MILAN,ITALY
关键词
HEAT AND MOISTURE EXCHANGE FILTERS; AIRWAY HUMIDIFICATION;
D O I
10.1007/BF01694711
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To evaluate the effect of different room temperatures on hydrophobic heat and moisture exchangers (HME) humidifying capability and efficiency. Methods: Stock HMEs were tested in vitro using an already described test model, with separation of inspiratory and expiratory gas. Absolute humidity (AH) was measured by means of dry-wet dual thermocouple, and HME efficiency was computed as the ratio between expired to inspired AH, at room temperature of 20 and 26-degrees-C. Results: Inspired gas temperature and AH were significantly higher at 26 than at 20-degrees-C; since expired AH remained substantially unchanged, HME efficiency was also higher in warmer environment. Conclusions: Hydrophobic HMEs appear to be affected by room temperature, increasing their humidifying ability and their efficiency with its rise.